Performance evaluation and analysis of CRP tested by nephelometry

Cui We

Open publisher page 0 citations

Abstract

Objective To evaluate the performance of CRP dectected by DADE BNII automatic special protein auto-analyzer.Methods The precision,accuracy,comparability and interference analysis were performed according to CLSI protocols(EP15-A,EP9-A2,EP7-A2) and other literatures.The results were compared with manufacturer's statement or international standards.Results The intra-assay inaccuracy(CV) of CRP at 13.84 mg/L and 55.6 mg/L was 3.32% and 3.58% respectively,which was less than the CV provided by manufacture(4.0% and 5.7%).The relative bias between the results detected for calibrator at two levels and given value from SIEMENS was 6.06% and 6.16%,less than 1/2 CLIA'88 standard(12.5%).The nephelometry was correlated with Roche Modular automatic biochemical analyzer(r=0.999 5,y=0.952 7x+0.355 2).The bias was acceptable between the two systems.Interference analysis showed the interference was limited when F-Bil at 0.2 g/L,hemoglobin at 4.82 g/L,chyle at 835 FTU.Conclusion The major performances of CRP detected by DADE BNII automatic special protein autoanlayzer was according with acceptable standards.

About this research paper

What this paper is about

Objective To evaluate the performance of CRP dectected by DADE BNII automatic special protein auto-analyzer.Methods The precision,accuracy,comparability and interference analysis were performed according to CLSI protocols(EP15-A,EP9-A2,EP7-A2) and other literatures.The results were compared with manufacturer's statement or international standards.Results The intra-assay inaccuracy(CV) of CRP at 13.84 mg/L and 55.6 mg/L was 3.32% and 3.58% respectively,which was less than the CV provided by manufacture(4.0% and 5.7%).The relative bias between the results detected for calibrator at two levels and given value from SIEMENS was 6.06% and 6.16%,less than 1/2 CLIA'88 standard(12.5%).The nephelometry was correlated with Roche Modular automatic biochemical analyzer(r=0.999 5,y=0.952 7x+0.355 2).The bias was acceptable between the two systems.Interference analysis showed the interference was limited when F-Bil at 0.2 g/L,hemoglobin at 4.82 g/L,chyle at 835 FTU.Conclusion The major performances of CRP detected by DADE BNII automatic special protein autoanlayzer was according with acceptable standards.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the performance of CRP dectected by DADE BNII automatic special protein auto-analyzer.Methods The precision,accuracy,comparability and interference analysis were performed according to CLSI protocols(EP15-A,EP9-A2,EP7-A2) and other literatures.The results were compared with manufacturer's statement or international standards.Results The intra-assay inaccuracy(CV) of CRP at 13.84 mg/L and 55.6 mg/L was 3.32% and 3.58% respectively,which was less than the CV provided by manufacture(4.0% and 5.7%).The relative bias between the results detected for calibrator at two levels and given value from SIEMENS was 6.06% and 6.16%,less than 1/2 CLIA'88 standard(12.5%).The nephelometry was correlated with Roche Modular automatic biochemical analyzer(r=0.999 5,y=0.952 7x+0.355 2).The bias was acceptable between the two systems.Interference analysis showed the interference was limited when F-Bil at 0.2 g/L,hemoglobin at 4.82 g/L,chyle at 835 FTU.Conclusion The major performances of CRP detected by DADE BNII automatic special protein autoanlayzer was according with acceptable standards.

Key concepts: Medicine, Nephelometry, Spectrum analyzer, Immunology, Optics, Physics, Antibody