The clinical reportable range of D-dimer on the automatic hemagglutinin apparatus Sysmex Ca1500
Jingjian Wang
Abstract
Jingjian Wang
Abstract
【Objective】To establish the clinical reportable range of D-dimer by immunoturbidimetry method.【Methods】The analytical measurement range(AMR) and the maximum allowable dilution rate were validated on Automatic Sysmex Ca1500,according to NCCLS EP6-P2 evaluation protocols and relevant literature.Combined with functional sensitivity,the clinical reportable range was established.【Results】 The AMR of D-dimer was during 29~1 980 ng/mL,and the maximum allowable dilution was 1 ∶ 8,which was consistent with the manufacturer's data.The clinical reportable range was 54.7~15 840 ng/mL.【Conclusions】 The clinical reportable range was established,which was significant for clinical diagnosis with the high test value.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
【Objective】To establish the clinical reportable range of D-dimer by immunoturbidimetry method.【Methods】The analytical measurement range(AMR) and the maximum allowable dilution rate were validated on Automatic Sysmex Ca1500,according to NCCLS EP6-P2 evaluation protocols and relevant literature.Combined with functional sensitivity,the clinical reportable range was established.【Results】 The AMR of D-dimer was during 29~1 980 ng/mL,and the maximum allowable dilution was 1 ∶ 8,which was consistent with the manufacturer's data.The clinical reportable range was 54.7~15 840 ng/mL.【Conclusions】 The clinical reportable range was established,which was significant for clinical diagnosis with the high test value.
Key concepts: Medicine, D-dimer, Clinical diagnosis, Internal medicine, Intensive care medicine