Performance Evaluation of Automatic Biochemical Analyzer in D-Dimer Test
Lili Li
Abstract
Lili Li
Abstract
Objective To evaluate the performance feature of Hitachi 7170 automatic biochemical analyzer in D-dimer test.Methods According to the evaluation protocols documents and peninent litemture,and by combining our actual work,the precision,accuracy,analyticai sensitivity,analytical measurement range,and the biological reference range of D-dimer were verified and evaluated.Results The inexactitude within-batch and day was superior to the standard of manufacturer.The related biases between results measured for calibrator at four levels and target valule was less 15%.Functional sensitivity was 0.46 μg/mL.Analytical measurement range was 0.30-42.95 μg/mL,that lower t than the manufacturer's claim slightly.The biological reference range was1 μg/mL,within manufacturer's claimed.Different concentrations of free bilirubin,direct bilirubin,chylomicrons,hemoglobin will not influence D-Dimer detection.Conclusion The main analytical performances of D-Dimer test with automatic biochemical analyzer are consistent with the claim defined by manufacturer,and could be satisfied in the clinical determination.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the performance feature of Hitachi 7170 automatic biochemical analyzer in D-dimer test.Methods According to the evaluation protocols documents and peninent litemture,and by combining our actual work,the precision,accuracy,analyticai sensitivity,analytical measurement range,and the biological reference range of D-dimer were verified and evaluated.Results The inexactitude within-batch and day was superior to the standard of manufacturer.The related biases between results measured for calibrator at four levels and target valule was less 15%.Functional sensitivity was 0.46 μg/mL.Analytical measurement range was 0.30-42.95 μg/mL,that lower t than the manufacturer's claim slightly.The biological reference range was1 μg/mL,within manufacturer's claimed.Different concentrations of free bilirubin,direct bilirubin,chylomicrons,hemoglobin will not influence D-Dimer detection.Conclusion The main analytical performances of D-Dimer test with automatic biochemical analyzer are consistent with the claim defined by manufacturer,and could be satisfied in the clinical determination.
Key concepts: Spectrum analyzer, D-dimer, Dimer, Accuracy and precision, Bilirubin, Hematology analyzer, Reference range, Chemistry