Initial evaluation of the ACL-Futura coagulation analyzer
Peng Chang-hu
Abstract
Peng Chang-hu
Abstract
Objective: To evaluate the technical performance of the ACL-Futura coagulation analyzer.And to provide data for various laboratories in the selection of coagulation analyzer.Methods: To co-evaluated the following assay:Prothrombin time(PT),Activated partial thromboplaslin time(APTT),Fibrinogen-C(Fbg-C,Clauss method),Fibrinogen-D(Fbg-D,Derived method) and Glycated Hemoglobin A1c(GHbA1c,Lactic-gel agglomerate method).The precision,stability,accuracy,linearity and interferences were assessed.Results:The analyzer shows an within-between day imprecision with normal control plasma,for PT of 1.0 to 1.3%,APTT of 1.8 to 3.8%,Fibrinogen-C of 2.4 to 3.3%,Fibrinogen-D of 1.4 to 1.5% and GHbA1c of 4.4 to 7.6% respectively.The calibration curves were linear within the concentration range of 1.75-5.31mg/L for Fbg-C,0.67-2.67mg/L for Fbg-D and 1.44-14.4 for GHbA1c(r0.998,P0.01).In addition,the correlation between the dilution of sample and the result(second) of APTT and PT is good(n=6,r0.999,P0.01).Experimental results also show that relative deviations of ATTT,PT,Fbg-C and Fbg-D were 3.9%,0.6%.0.6% and 0.4% respectively.It is not significantly affected by such biological products as haemoglobin,bilirubin and triglycerides up to 4.0g/L,127.3μmol/L and 6.36mmol/L respectively.Conclusion: Instrument is reliable,precision,easy to use and capable of fast sample throughput.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective: To evaluate the technical performance of the ACL-Futura coagulation analyzer.And to provide data for various laboratories in the selection of coagulation analyzer.Methods: To co-evaluated the following assay:Prothrombin time(PT),Activated partial thromboplaslin time(APTT),Fibrinogen-C(Fbg-C,Clauss method),Fibrinogen-D(Fbg-D,Derived method) and Glycated Hemoglobin A1c(GHbA1c,Lactic-gel agglomerate method).The precision,stability,accuracy,linearity and interferences were assessed.Results:The analyzer shows an within-between day imprecision with normal control plasma,for PT of 1.0 to 1.3%,APTT of 1.8 to 3.8%,Fibrinogen-C of 2.4 to 3.3%,Fibrinogen-D of 1.4 to 1.5% and GHbA1c of 4.4 to 7.6% respectively.The calibration curves were linear within the concentration range of 1.75-5.31mg/L for Fbg-C,0.67-2.67mg/L for Fbg-D and 1.44-14.4 for GHbA1c(r0.998,P0.01).In addition,the correlation between the dilution of sample and the result(second) of APTT and PT is good(n=6,r0.999,P0.01).Experimental results also show that relative deviations of ATTT,PT,Fbg-C and Fbg-D were 3.9%,0.6%.0.6% and 0.4% respectively.It is not significantly affected by such biological products as haemoglobin,bilirubin and triglycerides up to 4.0g/L,127.3μmol/L and 6.36mmol/L respectively.Conclusion: Instrument is reliable,precision,easy to use and capable of fast sample throughput.
Key concepts: Spectrum analyzer, Chromatography, Fibrinogen, Prothrombin time, Coagulation, Calibration curve, Chemistry, Analytical Chemistry (journal)