Performance Verification of ACL-TOP Automatic Coagulation Analyzer Detection System
Zhu Mei-cai
Abstract
Zhu Mei-cai
Abstract
Objective To verify the main performances of ACL-TOP automatic coagulation analyzer from Beckman-Coulter of America.Methods Prothrombin time(PT),active partial thromboplastin time(APTT),fibrinogen(FIB),thrombin time(TT) and D-Dimer(D-D) were selected to verify the performance of the equipment.The precision was evaluated by coefficient variation(CV),carry-over by carry-over rate,linear range by regression equation and the comparison of different machines by relative deviation.Results The precision of PT,APTT,FIB,TT and D-D were lower than 6%.All the precision of selected testing items were within an acceptable range.The carryover rates of PT,APTT,FIB,TT and D-D were 0.43%,1.56%,1.06%,1.43% and 1.98%,respectively.The linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.97~1.03,and r0.975.The relative deviation in different machines was lower than 1/2 total errors(TE) provided by equipment factory.Conclusion The precision,carryover rates,the comparison results in different machines and linearity of the selected testing items of ACL-TOP automatic coagulation analyzer accord with the demands of the experiments in clinical diagnosis,which can be used clinically.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To verify the main performances of ACL-TOP automatic coagulation analyzer from Beckman-Coulter of America.Methods Prothrombin time(PT),active partial thromboplastin time(APTT),fibrinogen(FIB),thrombin time(TT) and D-Dimer(D-D) were selected to verify the performance of the equipment.The precision was evaluated by coefficient variation(CV),carry-over by carry-over rate,linear range by regression equation and the comparison of different machines by relative deviation.Results The precision of PT,APTT,FIB,TT and D-D were lower than 6%.All the precision of selected testing items were within an acceptable range.The carryover rates of PT,APTT,FIB,TT and D-D were 0.43%,1.56%,1.06%,1.43% and 1.98%,respectively.The linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.97~1.03,and r0.975.The relative deviation in different machines was lower than 1/2 total errors(TE) provided by equipment factory.Conclusion The precision,carryover rates,the comparison results in different machines and linearity of the selected testing items of ACL-TOP automatic coagulation analyzer accord with the demands of the experiments in clinical diagnosis,which can be used clinically.
Key concepts: Spectrum analyzer, Prothrombin time, Coefficient of variation, Linear regression, Mathematics, Linearity, Fibrinogen, Partial thromboplastin time