Performance Verification of ACL-TOP Automatic Coagulation Analyzer Detection System

Zhu Mei-cai

Open publisher page 0 citations

Abstract

Objective To verify the main performances of ACL-TOP automatic coagulation analyzer from Beckman-Coulter of America.Methods Prothrombin time(PT),active partial thromboplastin time(APTT),fibrinogen(FIB),thrombin time(TT) and D-Dimer(D-D) were selected to verify the performance of the equipment.The precision was evaluated by coefficient variation(CV),carry-over by carry-over rate,linear range by regression equation and the comparison of different machines by relative deviation.Results The precision of PT,APTT,FIB,TT and D-D were lower than 6%.All the precision of selected testing items were within an acceptable range.The carryover rates of PT,APTT,FIB,TT and D-D were 0.43%,1.56%,1.06%,1.43% and 1.98%,respectively.The linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.97~1.03,and r0.975.The relative deviation in different machines was lower than 1/2 total errors(TE) provided by equipment factory.Conclusion The precision,carryover rates,the comparison results in different machines and linearity of the selected testing items of ACL-TOP automatic coagulation analyzer accord with the demands of the experiments in clinical diagnosis,which can be used clinically.

About this research paper

What this paper is about

Objective To verify the main performances of ACL-TOP automatic coagulation analyzer from Beckman-Coulter of America.Methods Prothrombin time(PT),active partial thromboplastin time(APTT),fibrinogen(FIB),thrombin time(TT) and D-Dimer(D-D) were selected to verify the performance of the equipment.The precision was evaluated by coefficient variation(CV),carry-over by carry-over rate,linear range by regression equation and the comparison of different machines by relative deviation.Results The precision of PT,APTT,FIB,TT and D-D were lower than 6%.All the precision of selected testing items were within an acceptable range.The carryover rates of PT,APTT,FIB,TT and D-D were 0.43%,1.56%,1.06%,1.43% and 1.98%,respectively.The linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.97~1.03,and r0.975.The relative deviation in different machines was lower than 1/2 total errors(TE) provided by equipment factory.Conclusion The precision,carryover rates,the comparison results in different machines and linearity of the selected testing items of ACL-TOP automatic coagulation analyzer accord with the demands of the experiments in clinical diagnosis,which can be used clinically.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To verify the main performances of ACL-TOP automatic coagulation analyzer from Beckman-Coulter of America.Methods Prothrombin time(PT),active partial thromboplastin time(APTT),fibrinogen(FIB),thrombin time(TT) and D-Dimer(D-D) were selected to verify the performance of the equipment.The precision was evaluated by coefficient variation(CV),carry-over by carry-over rate,linear range by regression equation and the comparison of different machines by relative deviation.Results The precision of PT,APTT,FIB,TT and D-D were lower than 6%.All the precision of selected testing items were within an acceptable range.The carryover rates of PT,APTT,FIB,TT and D-D were 0.43%,1.56%,1.06%,1.43% and 1.98%,respectively.The linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.97~1.03,and r0.975.The relative deviation in different machines was lower than 1/2 total errors(TE) provided by equipment factory.Conclusion The precision,carryover rates,the comparison results in different machines and linearity of the selected testing items of ACL-TOP automatic coagulation analyzer accord with the demands of the experiments in clinical diagnosis,which can be used clinically.

Key concepts: Spectrum analyzer, Prothrombin time, Coefficient of variation, Linear regression, Mathematics, Linearity, Fibrinogen, Partial thromboplastin time

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance Verification of ACL-TOP Automatic Coagulation Analyzer Detection System — Research Paper | ScholarLens