Performance Evaluation of Sysmex XS-800i Hematology Analyzer

Chengbin Wang

Open publisher page 0 citations

Abstract

Objective To evaluate the performance of Sysmex XS-800i hematology analyzer on CBC and DC. Methods The indexes including precision, total reproducibility, carry-over, linearity (dilution effect) and comparability were examined according to the relevant standard and compares with manual classification. Results The CBC of precision, carry-over rate and total reproducibility was meet the request. The instrument was better than XE-2100 in correlation(r0.99). Leukocyte differential count: Neut, Lym, Mon, Eos and Baso has good reproducibility. Among the comparisons of normal samples and microscopic effect, the Neut, Lym and Eos first place(r=0.9816, 0.9795 and 0.9629), the Mon second(r=0.828 3), the Baso comes last (r=0.587 4). Conclusion The analyzer can be used to test blood cell parameters accurately and reli-ably. Its main performance index meets the experimental requirements. It must be approved by the microscope for abnor-mal example of WBC before report.

About this research paper

What this paper is about

Objective To evaluate the performance of Sysmex XS-800i hematology analyzer on CBC and DC. Methods The indexes including precision, total reproducibility, carry-over, linearity (dilution effect) and comparability were examined according to the relevant standard and compares with manual classification. Results The CBC of precision, carry-over rate and total reproducibility was meet the request. The instrument was better than XE-2100 in correlation(r0.99). Leukocyte differential count: Neut, Lym, Mon, Eos and Baso has good reproducibility. Among the comparisons of normal samples and microscopic effect, the Neut, Lym and Eos first place(r=0.9816, 0.9795 and 0.9629), the Mon second(r=0.828 3), the Baso comes last (r=0.587 4). Conclusion The analyzer can be used to test blood cell parameters accurately and reli-ably. Its main performance index meets the experimental requirements. It must be approved by the microscope for abnor-mal example of WBC before report.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the performance of Sysmex XS-800i hematology analyzer on CBC and DC. Methods The indexes including precision, total reproducibility, carry-over, linearity (dilution effect) and comparability were examined according to the relevant standard and compares with manual classification. Results The CBC of precision, carry-over rate and total reproducibility was meet the request. The instrument was better than XE-2100 in correlation(r0.99). Leukocyte differential count: Neut, Lym, Mon, Eos and Baso has good reproducibility. Among the comparisons of normal samples and microscopic effect, the Neut, Lym and Eos first place(r=0.9816, 0.9795 and 0.9629), the Mon second(r=0.828 3), the Baso comes last (r=0.587 4). Conclusion The analyzer can be used to test blood cell parameters accurately and reli-ably. Its main performance index meets the experimental requirements. It must be approved by the microscope for abnor-mal example of WBC before report.

Key concepts: Hematology analyzer, Reproducibility, Medicine, Spectrum analyzer, Nuclear medicine, Internal medicine, Medical physics, Mathematics

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance Evaluation of Sysmex XS-800i Hematology Analyzer — Research Paper | ScholarLens