Aim To establish an HPLC method for determination of matrine in Xiaomishuan.Methods The LabAlliance C18 column with methanol-water-triethylamine(45:55:0.275)was used as the mobile phase.The detection flow was 1.0 ml·min-1.The detection wavelength was at 220 nm.Results The standard curves were in good linear relationship at the rang e from 80 to 400 mg·L-1,r=0.9995.The average recovery was 98.7%(n=6)with RSD of 1.22%.Conclusions The method is simple and accurate with good reproducibility and can be used for the quality control.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Aim To establish an HPLC method for determination of matrine in Xiaomishuan.Methods The LabAlliance C18 column with methanol-water-triethylamine(45:55:0.275)was used as the mobile phase.The detection flow was 1.0 ml·min-1.The detection wavelength was at 220 nm.Results The standard curves were in good linear relationship at the rang e from 80 to 400 mg·L-1,r=0.9995.The average recovery was 98.7%(n=6)with RSD of 1.22%.Conclusions The method is simple and accurate with good reproducibility and can be used for the quality control.