Study on Corneal thickness measurements with the Orbscan-II Topography and ultrasonic pachymetry before and after LASIK
Mei Hui-xiang
Abstract
Mei Hui-xiang
Abstract
Objective To investigate the difference on corneal thickness measurements using Orbscan-ⅡSystem and ultrasonic pachymetry before and after LASIK in myopic patients.Methods Orbscan-ⅡSystem and A-scan ultrasonography were used to measure the corneal thickness pre-and post-operative in 300 eyes of 150 cases that underwent myopic LASIK.The differences were evaluated by t tests.Results Pre-operatively,the mean corneal thickness was 535.17±32.352 μm with Orbscan-Ⅱsystem,and 545.43±29.079 μm with A-scan ultrasonography.Post-operatively,the mean corneal thickness with Orbscan-Ⅱsystem and with A-scan ultrasonography was 417.06±46.422 μm and 457.64±32.30 μm respectively,and the differences were statistically significant.Conclusion There is significant difference between the corneal thickness value measured by Orbscan-II topography system and ultrasonic pachymetry.Even high correlation between them has been showed,Orbscan-Ⅱsystem for corneal thickness measurement could not substitute A-scan ultrasonography completely.Certainly it still can be used as the very good reference for LASIK.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To investigate the difference on corneal thickness measurements using Orbscan-ⅡSystem and ultrasonic pachymetry before and after LASIK in myopic patients.Methods Orbscan-ⅡSystem and A-scan ultrasonography were used to measure the corneal thickness pre-and post-operative in 300 eyes of 150 cases that underwent myopic LASIK.The differences were evaluated by t tests.Results Pre-operatively,the mean corneal thickness was 535.17±32.352 μm with Orbscan-Ⅱsystem,and 545.43±29.079 μm with A-scan ultrasonography.Post-operatively,the mean corneal thickness with Orbscan-Ⅱsystem and with A-scan ultrasonography was 417.06±46.422 μm and 457.64±32.30 μm respectively,and the differences were statistically significant.Conclusion There is significant difference between the corneal thickness value measured by Orbscan-II topography system and ultrasonic pachymetry.Even high correlation between them has been showed,Orbscan-Ⅱsystem for corneal thickness measurement could not substitute A-scan ultrasonography completely.Certainly it still can be used as the very good reference for LASIK.
Key concepts: Medicine, Ophthalmology, LASIK, Ultrasonography, Ultrasonic sensor, Corneal topography, Significant difference, Cornea