CLINICAL APPLICATION OF AUTO-MILLIAMPERE TECHNIQUE OF SPIRAL CT
Zhou Huang
Abstract
Zhou Huang
Abstract
[Objective]This study aimed to explore the changes of noise figure,radiation dose accepted by the patients and image quality with the application of automatic mA technology of whole abdominal CT scanning.[Methods]40 patients of adult full abdominal examination were randomly divided into 4 groups,A group used the conventional abdominal scan parameters as a reference group,B,C,D group used the automatic mA scanning technology;the average mAs,CTDIvol and DLP were recorded in each group.[Results]The radiation dose under auto mA scan decreased significantly compared with conventional parameters.The noise figure was higher,the scanning dose was lower,then image quality reduced.[Conclusion]The appropriate noise index mA under whole abdominal CT scan can automatically ensure image quality,but also significantly reduce the acceptable radiation dose in patients.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
[Objective]This study aimed to explore the changes of noise figure,radiation dose accepted by the patients and image quality with the application of automatic mA technology of whole abdominal CT scanning.[Methods]40 patients of adult full abdominal examination were randomly divided into 4 groups,A group used the conventional abdominal scan parameters as a reference group,B,C,D group used the automatic mA scanning technology;the average mAs,CTDIvol and DLP were recorded in each group.[Results]The radiation dose under auto mA scan decreased significantly compared with conventional parameters.The noise figure was higher,the scanning dose was lower,then image quality reduced.[Conclusion]The appropriate noise index mA under whole abdominal CT scan can automatically ensure image quality,but also significantly reduce the acceptable radiation dose in patients.
Key concepts: Image quality, Medicine, Image noise, Radiation dose, Nuclear medicine, Spiral (railway), Radiation exposure, Noise (video)