Performance Evaluation of LH 750 Automatic Hematology Analyzer

Hong Zhu

Open publisher page 0 citations

Abstract

Objective To evaluate the performance of LH 750 hematology analyzer.Methods The indexes including precision,total reproducibility,carry-over,linearity and comparability were examined according to published evaluation standard.At the same time,the indexes of differential count(DC) were compared between those produced by the analyzer and manual methods.Results Precision,carry-over and total reproducibility in counting of leucocytes,erythrocytes,platelets and examination of hemoglobin on the analyzer were showed in the prescriptive range.The determined results for linearity-control were expected ranges.The results correlated well(r20.963) and bias were little(-2.094%~1.284%) with those obtained on the comparative instrument.DC:reproducibility was good for neutrophils,lymphocytes,monocytes,eosinophils and basophils.Comparison of the results by instrument with manual for normal samples in morphology,the correlation was better for neutrophils,lymphocytes and eosinophils(r2=0.946 9~0.966 1);good for monocytes(r2=0.801 3),not good for basophils(r2=0.518 0).Comparison of instrument flags with manual DC for 296 samples,false positive ratio,false negative ratio,accuracy,specificity and sensitivity were 15.93%,4.35%,86.78%,84.07% and 95.65% separately.Conclusion The analyzer can be used to test blood cell parameters accurately and reliably.Its main performance indices accords with the experimental requirements;the results are credible.It is necessary to check with microscopy for DC before reported when it is doubted.

About this research paper

What this paper is about

Objective To evaluate the performance of LH 750 hematology analyzer.Methods The indexes including precision,total reproducibility,carry-over,linearity and comparability were examined according to published evaluation standard.At the same time,the indexes of differential count(DC) were compared between those produced by the analyzer and manual methods.Results Precision,carry-over and total reproducibility in counting of leucocytes,erythrocytes,platelets and examination of hemoglobin on the analyzer were showed in the prescriptive range.The determined results for linearity-control were expected ranges.The results correlated well(r20.963) and bias were little(-2.094%~1.284%) with those obtained on the comparative instrument.DC:reproducibility was good for neutrophils,lymphocytes,monocytes,eosinophils and basophils.Comparison of the results by instrument with manual for normal samples in morphology,the correlation was better for neutrophils,lymphocytes and eosinophils(r2=0.946 9~0.966 1);good for monocytes(r2=0.801 3),not good for basophils(r2=0.518 0).Comparison of instrument flags with manual DC for 296 samples,false positive ratio,false negative ratio,accuracy,specificity and sensitivity were 15.93%,4.35%,86.78%,84.07% and 95.65% separately.Conclusion The analyzer can be used to test blood cell parameters accurately and reliably.Its main performance indices accords with the experimental requirements;the results are credible.It is necessary to check with microscopy for DC before reported when it is doubted.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the performance of LH 750 hematology analyzer.Methods The indexes including precision,total reproducibility,carry-over,linearity and comparability were examined according to published evaluation standard.At the same time,the indexes of differential count(DC) were compared between those produced by the analyzer and manual methods.Results Precision,carry-over and total reproducibility in counting of leucocytes,erythrocytes,platelets and examination of hemoglobin on the analyzer were showed in the prescriptive range.The determined results for linearity-control were expected ranges.The results correlated well(r20.963) and bias were little(-2.094%~1.284%) with those obtained on the comparative instrument.DC:reproducibility was good for neutrophils,lymphocytes,monocytes,eosinophils and basophils.Comparison of the results by instrument with manual for normal samples in morphology,the correlation was better for neutrophils,lymphocytes and eosinophils(r2=0.946 9~0.966 1);good for monocytes(r2=0.801 3),not good for basophils(r2=0.518 0).Comparison of instrument flags with manual DC for 296 samples,false positive ratio,false negative ratio,accuracy,specificity and sensitivity were 15.93%,4.35%,86.78%,84.07% and 95.65% separately.Conclusion The analyzer can be used to test blood cell parameters accurately and reliably.Its main performance indices accords with the experimental requirements;the results are credible.It is necessary to check with microscopy for DC before reported when it is doubted.

Key concepts: Hematology analyzer, Reproducibility, Spectrum analyzer, Medicine, Biomedical engineering, Linearity, Mathematics, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance Evaluation of LH 750 Automatic Hematology Analyzer — Research Paper | ScholarLens