Multicentre evaluation of a second generation point-of-care assay with an extended range for the determination of N-terminal pro-brain natriuretic peptide.
Bo Jørgensen, Thomas Bertsch, Hans-Joachim Bröker, Markus Schäfer, Jean-Paul Chapelle, Romy Gadisseur, Martín Cowie, Bert D. Dikkeschei, Eberhard Gurr, Wiebke Hayen, Gerd Häfner, Yuriko Stiegler, Hans‐Robert Schönherr, Ruth H. Strasser, Britta Weidtmann, Henning Folkerts, Christian Zugck, Kerstin Hofmann, Rainer Zerback
Abstract
BACKGROUND: In the second generation of the point-of-care (POC) assay Roche CARDIAC proBNP, the upper limit of the measuring range was extended from 3000 to 9000 ng/L. METHODS: A thirteen-site multicentre evaluation was carried out to assess the analytical performance of the POC NT-proBNP assay and to compare it with a laboratory N-terminal pro-brain natriuretic peptide (NT-proBNP) assay. RESULTS: In method comparisons of six lots of POC NT-proBNP with the lab reference method (Elecsys proBNP) mean bias ranged from -10 to +17%. In lot-to-lot comparisons all six investigated lots of POC NT-proBNP showed excellent agreement, with mean bias between -7% and +2%. The majority of all coefficients of variation obtained from ten-fold measurements using 56 native blood samples were below 8%. No interference was observed with hemolytic blood (hemoglobin concentrations up to 0.12 mmol/L), lipemic blood (triglyceride concentrations up to 14.0 mmol/L) nor icteric blood (bilirubin concentrations up to 63 micromol/L). Hematocrit values between 24% and 51% had no influence on the assay result. High NT-proBNP concentrations above the measuring range of POC NT-proBNP did not lead to false low results due to potential high-dose hook effect. Results with POC NT-proBNP were not influenced by different ambient temperatures (18 degrees C to 32 degrees C), the sample material used, nor by over- or underdosing by 15 microL compared to the regular sample volume of 150 microL. CONCLUSIONS: The POC NT-proBNP assay showed an excellent analytical performance including a good agreement with the laboratory method. The assay is therefore suitable for its intended use in point-of-care settings.