Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods.
Sidney M. Hopfer, Francesca L. Nadeau, Marilyn Sundra, Gregory S. Makowski
Abstract
Sidney M. Hopfer, Francesca L. Nadeau, Marilyn Sundra, Gregory S. Makowski
Abstract
Point-of-care testing (POCT) for blood hemoglobin and hematocrit (H/H) levels provides rapid patient assessment including the need for transfusion. Conductivity-based methods of blood H/H determinations can be influenced by plasma protein concentration. To assess this factor, we measured H/H levels at varying protein concentrations using two POCT instruments: iSTAT-1 (conductivity method) and Hemocue (optical method). These H/H results were compared to results obtained by our core laboratory hematology analyzer (GenS). Anticoagulated whole blood was centrifuged to sediment the red blood cells; the plasma was removed to serve as source of protein for mixing studies. A series of reconstituted samples was prepared with varying H/H and protein levels. To mimic hemodilution in clinical practice, samples were diluted with saline or lactated Ringer's solutions. Following H/H analysis, the samples were centrifuged and protein determined in the supernatant plasmas. The H/H results obtained with the Hemocue instrument correlated exactly with those of the GenS analyzer at protein concentrations from 0.7 to 6.2 g/dl. The correlation was unaffected when hemodilution was performed with either saline (r = 0.999) or lactated Ringer's (r = 1.000). The H/H results obtained with the iSTAT-1 instrument gave slightly less correlation with those of the GenS analyzer (r = 0.978 - 0.980) over this protein range. However, the iSTAT-1 results were generally lower than the GenS results, with discrepancies up to 2 g/dL for hemoglobin values and up to 4% for hematocrits at the lowest protein concentration. Therefore, it is recommended that H/H testing in patients with suspected hypoproteinemia or substantial hemodilution should be tested with a non-conductivity-based method.
OpenAlex reports 50 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Point-of-care testing (POCT) for blood hemoglobin and hematocrit (H/H) levels provides rapid patient assessment including the need for transfusion. Conductivity-based methods of blood H/H determinations can be influenced by plasma protein concentration. To assess this factor, we measured H/H levels at varying protein concentrations using two POCT instruments: iSTAT-1 (conductivity method) and Hemocue (optical method). These H/H results were compared to results obtained by our core laboratory hematology analyzer (GenS). Anticoagulated whole blood was centrifuged to sediment the red blood cells; the plasma was removed to serve as source of protein for mixing studies. A series of reconstituted samples was prepared with varying H/H and protein levels. To mimic hemodilution in clinical practice, samples were diluted with saline or lactated Ringer's solutions. Following H/H analysis, the samples were centrifuged and protein determined in the supernatant plasmas. The H/H results obtained with the Hemocue instrument correlated exactly with those of the GenS analyzer at protein concentrations from 0.7 to 6.2 g/dl. The correlation was unaffected when hemodilution was performed with either saline (r = 0.999) or lactated Ringer's (r = 1.000). The H/H results obtained with the iSTAT-1 instrument gave slightly less correlation with those of the GenS analyzer (r = 0.978 - 0.980) over this protein range. However, the iSTAT-1 results were generally lower than the GenS results, with discrepancies up to 2 g/dL for hemoglobin values and up to 4% for hematocrits at the lowest protein concentration. Therefore, it is recommended that H/H testing in patients with suspected hypoproteinemia or substantial hemodilution should be tested with a non-conductivity-based method.
Key concepts: Hematocrit, Point-of-care testing, Hemoglobin, Saline, Chromatography, Chemistry, Whole blood, Point of care