Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer

Tarek Saleh, M K M Adams, Bill McDermott, Kate G Claridge, Paul Ewings

Open full text 30 citations

Abstract

PURPOSE: To assess the effects of central corneal thickness (CCT) and corneal curvature (CC) on the measurements of intraocular pressure (IOP) using Goldmann applanation tonometer (GAT) and the ocular blood flow pneumatonometer (OBFT). METHODS: 104 patients were recruited from a glaucoma clinic. The CCT was measured using ultrasound pachymetry and the mean radius of CC using a keratometer. The IOP of each eye was measured using both GAT and the OBFT in a random order. Right eyes only were analysed for statistical purposes. RESULTS: The mean (+/-SD) IOP by GAT and OBFT was 18.2 mmHg (+/-4.4) and 18.2 mmHg (+/-4.0), respectively, with no statistically significant difference. IOP measurement with both instruments varied with CCT and CC. GAT showed an IOP increase of 0.40 mmHg per 10 microm increase of CCT and OBFT showed an increase of 0.38 mmHg in IOP per 10 microm increase of CCT. Multiple regression analysis showed that the effect of CCT was statistically significant (P<0.001) on IOP recorded by both the GAT and OBFT but CC did not have a statistically significant effect on IOP recordings performed by either technique. CONCLUSION: IOP measurements by GAT and OBFT are positively correlated with CCT with both tonometers being similarly affected. There was no significant correlation between CC and IOP measured by either tonometer.

Open-access reader

About this research paper

What this paper is about

PURPOSE: To assess the effects of central corneal thickness (CCT) and corneal curvature (CC) on the measurements of intraocular pressure (IOP) using Goldmann applanation tonometer (GAT) and the ocular blood flow pneumatonometer (OBFT). METHODS: 104 patients were recruited from a glaucoma clinic. The CCT was measured using ultrasound pachymetry and the mean radius of CC using a keratometer. The IOP of each eye was measured using both GAT and the OBFT in a random order. Right eyes only were analysed for statistical purposes. RESULTS: The mean (+/-SD) IOP by GAT and OBFT was 18.2 mmHg (+/-4.4) and 18.2 mmHg (+/-4.0), respectively, with no statistically significant difference. IOP measurement with both instruments varied with CCT and CC. GAT showed an IOP increase of 0.40 mmHg per 10 microm increase of CCT and OBFT showed an increase of 0.38 mmHg in IOP per 10 microm increase of CCT. Multiple regression analysis showed that the effect of CCT was statistically significant (P<0.001) on IOP recorded by both the GAT and OBFT but CC did not have a statistically significant effect on IOP recordings performed by either technique. CONCLUSION: IOP measurements by GAT and OBFT are positively correlated with CCT with both tonometers being similarly affected. There was no significant correlation between CC and IOP measured by either tonometer.

Why it matters

OpenAlex reports 30 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: To assess the effects of central corneal thickness (CCT) and corneal curvature (CC) on the measurements of intraocular pressure (IOP) using Goldmann applanation tonometer (GAT) and the ocular blood flow pneumatonometer (OBFT). METHODS: 104 patients were recruited from a glaucoma clinic. The CCT was measured using ultrasound pachymetry and the mean radius of CC using a keratometer. The IOP of each eye was measured using both GAT and the OBFT in a random order. Right eyes only were analysed for statistical purposes. RESULTS: The mean (+/-SD) IOP by GAT and OBFT was 18.2 mmHg (+/-4.4) and 18.2 mmHg (+/-4.0), respectively, with no statistically significant difference. IOP measurement with both instruments varied with CCT and CC. GAT showed an IOP increase of 0.40 mmHg per 10 microm increase of CCT and OBFT showed an increase of 0.38 mmHg in IOP per 10 microm increase of CCT. Multiple regression analysis showed that the effect of CCT was statistically significant (P<0.001) on IOP recorded by both the GAT and OBFT but CC did not have a statistically significant effect on IOP recordings performed by either technique. CONCLUSION: IOP measurements by GAT and OBFT are positively correlated with CCT with both tonometers being similarly affected. There was no significant correlation between CC and IOP measured by either tonometer.

Key concepts: Medicine, Ophthalmology, Intraocular pressure, Goldmann Applanation Tonometer, Keratometer, Glaucoma, Applanation tonometry, Cornea