Comparison of central corneal thickness measurements by Ultrasound,Orbscan II ,AS-OCT and Pentacam pachymetry
Xianming Jiang, Bin Yang, Zheng Wang, Guofu Huang, Xiaoxiao Zhang, Sheng Zhou
Abstract
Xianming Jiang, Bin Yang, Zheng Wang, Guofu Huang, Xiaoxiao Zhang, Sheng Zhou
Abstract
Objective To analyze the difference and the agreement in central corneal thickness(CCT) measurements between the gold standard method of ultrasound pachymetry and 3 noncontact systems(Orbscan Ⅱ, AS-OCT, Pentacam)in ametropia eyes before laser refractive surgery.Methods Ninety consecutive patients(180 eyes)who had preoperative examination in Laser Refractive Surgery Center had Orbscan Ⅱ, AS-OCT, Pentacam, and ultrasound pachymetry.Data were analyzed using the paired sample t test, linear regression, and Bland-Altman plots.Results The mean CCT measured by ultrasound,OrbscanⅡ,Pentacam,and AS-OCT pacbymetry were,respectively,(542.1 ±30.4)μ m、(541.1±37.1)μ m、(530.3±29.3)μ m and(544.9±8.7)μ m.There was no statistically significant difference between ultrasound and OrbscanⅡ measurements(P>0.05).Compared with the ultrasound measurement,AS-OCT and Pentacam measurements significantly misestimate corneal thickness by a mean of(11.7±8.9)μ m(P <0.01)and(-2.8±1.0)μ m(P <0.01),re-spectively.OrbscanⅡ, AS-OCT, Pentacam pachymetry showed a high correlation with ultrasound pachymetry. Tbe95%limitsofagreement(LoA)withultrasoundwere-37.4~+39.3 μ m for the OrbscanⅡdevice and-5.6~+29.1 μ m for the AS-OCT device and -24.4~+18.8 μ m for the Pentacam device.Conclusions OrbscanⅡ,AS-OCT and Pentacam pacbymetry can be used to measure central corneal thickness.Compared with the ultrasound pachymetry, OrbscanⅡ pachymetry was effectd by it's AF, AS-OCT pachymetry had the best agreement, Penta-cam pacbymetry had the nearest CCT.However it is cautious to use the 4 devices interchangeably. Key words: Ultrasound pachymeter; Orbscan Ⅱ ; AS-OCT; Pentacam; Agreement
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To analyze the difference and the agreement in central corneal thickness(CCT) measurements between the gold standard method of ultrasound pachymetry and 3 noncontact systems(Orbscan Ⅱ, AS-OCT, Pentacam)in ametropia eyes before laser refractive surgery.Methods Ninety consecutive patients(180 eyes)who had preoperative examination in Laser Refractive Surgery Center had Orbscan Ⅱ, AS-OCT, Pentacam, and ultrasound pachymetry.Data were analyzed using the paired sample t test, linear regression, and Bland-Altman plots.Results The mean CCT measured by ultrasound,OrbscanⅡ,Pentacam,and AS-OCT pacbymetry were,respectively,(542.1 ±30.4)μ m、(541.1±37.1)μ m、(530.3±29.3)μ m and(544.9±8.7)μ m.There was no statistically significant difference between ultrasound and OrbscanⅡ measurements(P>0.05).Compared with the ultrasound measurement,AS-OCT and Pentacam measurements significantly misestimate corneal thickness by a mean of(11.7±8.9)μ m(P <0.01)and(-2.8±1.0)μ m(P <0.01),re-spectively.OrbscanⅡ, AS-OCT, Pentacam pachymetry showed a high correlation with ultrasound pachymetry. Tbe95%limitsofagreement(LoA)withultrasoundwere-37.4~+39.3 μ m for the OrbscanⅡdevice and-5.6~+29.1 μ m for the AS-OCT device and -24.4~+18.8 μ m for the Pentacam device.Conclusions OrbscanⅡ,AS-OCT and Pentacam pacbymetry can be used to measure central corneal thickness.Compared with the ultrasound pachymetry, OrbscanⅡ pachymetry was effectd by it's AF, AS-OCT pachymetry had the best agreement, Penta-cam pacbymetry had the nearest CCT.However it is cautious to use the 4 devices interchangeably. Key words: Ultrasound pachymeter; Orbscan Ⅱ ; AS-OCT; Pentacam; Agreement
Key concepts: Medicine, Ophthalmology, Ultrasound, Significant difference, Refractive surgery, Optometry, Cornea, Radiology