Clinical Evaluation of New Office Spirometer
Leonard D. Hudson
Abstract
Leonard D. Hudson
Abstract
The accuracy of a simple, pneumatic, direct-recording spirometer suitable for office use was evaluated by comparing spirometry on a water-sealed, 13.5-liter, water-filled spirometer for 120 patients. Good correlation between the two spirometers was seen through a wide range of values for forced vital capacity, forced expiratory volume in one second, and forced expiratory flow during 25% to 75% of forced vital capacity, with coefficients of correlation being .988, .988, and .948, respectively. All correlations were significant. The pneumatic spirometer is accurate, simple to operate, and suitable for spirometry in the office and clinic.
OpenAlex reports 8 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The accuracy of a simple, pneumatic, direct-recording spirometer suitable for office use was evaluated by comparing spirometry on a water-sealed, 13.5-liter, water-filled spirometer for 120 patients. Good correlation between the two spirometers was seen through a wide range of values for forced vital capacity, forced expiratory volume in one second, and forced expiratory flow during 25% to 75% of forced vital capacity, with coefficients of correlation being .988, .988, and .948, respectively. All correlations were significant. The pneumatic spirometer is accurate, simple to operate, and suitable for spirometry in the office and clinic.
Key concepts: Spirometer, Vital capacity, Spirometry, Medicine, Physical therapy, Lung function, Asthma, Internal medicine