1979Applied OpticsRequires access

NBS reference hazemeter: its development and testing

Victor R. Weidner, Jack J. Hsia

Open publisher page 14 citations

Abstract

A research hazemeter has been developed by the Radiometric Physics Division of the National Bureau of Standards for the quantitative analysis of light scattering by optical materials. The instrument is designed to measure percentage haze as set forth in the prescribed conditions recommended in ASTM Test Method D 1003. The hazemeter has several special features which make it possible to vary the geometrical conditions of the collimated light beam and the integrating sphere detector system for detailed studies of light scattering. Light scattering associated with the instrument source and collimating optics has been reduced to less than 0.1% of the total light beam by using a diffraction limited off-axis parabolic mirror to collimate the hazemeter light beam. The source optics provide a collimated light beam of approximately 6800 K. The detector is a silicon cell photodiode with a V(lambda) window. This combination gives the system a response corresponding to the visual luminosity function. The detector nonlinearity and instrumental errors associated with the accuracy of the integrating sphere geometry have been analyzed. The over-all uncertainty in percentage haze has been reduced to +/-0.2%, and light scattering of less than 0.5% can readily be measured.

About this research paper

What this paper is about

A research hazemeter has been developed by the Radiometric Physics Division of the National Bureau of Standards for the quantitative analysis of light scattering by optical materials. The instrument is designed to measure percentage haze as set forth in the prescribed conditions recommended in ASTM Test Method D 1003. The hazemeter has several special features which make it possible to vary the geometrical conditions of the collimated light beam and the integrating sphere detector system for detailed studies of light scattering. Light scattering associated with the instrument source and collimating optics has been reduced to less than 0.1% of the total light beam by using a diffraction limited off-axis parabolic mirror to collimate the hazemeter light beam. The source optics provide a collimated light beam of approximately 6800 K. The detector is a silicon cell photodiode with a V(lambda) window. This combination gives the system a response corresponding to the visual luminosity function. The detector nonlinearity and instrumental errors associated with the accuracy of the integrating sphere geometry have been analyzed. The over-all uncertainty in percentage haze has been reduced to +/-0.2%, and light scattering of less than 0.5% can readily be measured.

Why it matters

OpenAlex reports 14 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A research hazemeter has been developed by the Radiometric Physics Division of the National Bureau of Standards for the quantitative analysis of light scattering by optical materials. The instrument is designed to measure percentage haze as set forth in the prescribed conditions recommended in ASTM Test Method D 1003. The hazemeter has several special features which make it possible to vary the geometrical conditions of the collimated light beam and the integrating sphere detector system for detailed studies of light scattering. Light scattering associated with the instrument source and collimating optics has been reduced to less than 0.1% of the total light beam by using a diffraction limited off-axis parabolic mirror to collimate the hazemeter light beam. The source optics provide a collimated light beam of approximately 6800 K. The detector is a silicon cell photodiode with a V(lambda) window. This combination gives the system a response corresponding to the visual luminosity function. The detector nonlinearity and instrumental errors associated with the accuracy of the integrating sphere geometry have been analyzed. The over-all uncertainty in percentage haze has been reduced to +/-0.2%, and light scattering of less than 0.5% can readily be measured.

Key concepts: Collimated light, Optics, Integrating sphere, Physics, Detector, Light scattering, Scattering, Photodiode

Related papers

Back to paper searchBrowse research topicsOriginal source
NBS reference hazemeter: its development and testing — Research Paper | ScholarLens