Improved control of image optical density with low-dose digital and conventional radiography in bedside imaging.

C. Schaefer, Reginald E. Greene, J W Oestmann, J M Kamalsky, D A Hall, HENRY J. LLEWELLYN, C L Robertson, James T. Rhea, HERBERT ROSENTHAL, JAMI R. RUBENS

Open publisher page 38 citations

Abstract

The technical and diagnostic performance of simultaneously acquired low-dose (44% of standard dose) storage-phosphor digital radiographs (system resolution = 0.2 mm, 10 bits) were compared with those of standard-dose conventional bedside radiographs of the chest in 32 patients. The mean optical density (OD) of the lungs (800 measurements) was closer to the ideal density with digital radiography (1.45 OD +/- 0.20 [standard deviation] vs 1.75 OD +/- 0.53) and was less often outside the usable range (2.5% vs 42.5%). Receiver operating characteristic analysis for detection of simulated nodules and monitoring devices (nine readers, 4,608 observations) showed that digital radiography was superior to conventional radiography (P less than .05) for four of the nine readers and equivalent to conventional radiography for five readers. The authors concluded that digital radiography produces more consistent and ideal image density and performs at least as well as conventional radiography under phantom test conditions.

About this research paper

What this paper is about

The technical and diagnostic performance of simultaneously acquired low-dose (44% of standard dose) storage-phosphor digital radiographs (system resolution = 0.2 mm, 10 bits) were compared with those of standard-dose conventional bedside radiographs of the chest in 32 patients. The mean optical density (OD) of the lungs (800 measurements) was closer to the ideal density with digital radiography (1.45 OD +/- 0.20 [standard deviation] vs 1.75 OD +/- 0.53) and was less often outside the usable range (2.5% vs 42.5%). Receiver operating characteristic analysis for detection of simulated nodules and monitoring devices (nine readers, 4,608 observations) showed that digital radiography was superior to conventional radiography (P less than .05) for four of the nine readers and equivalent to conventional radiography for five readers. The authors concluded that digital radiography produces more consistent and ideal image density and performs at least as well as conventional radiography under phantom test conditions.

Why it matters

OpenAlex reports 38 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The technical and diagnostic performance of simultaneously acquired low-dose (44% of standard dose) storage-phosphor digital radiographs (system resolution = 0.2 mm, 10 bits) were compared with those of standard-dose conventional bedside radiographs of the chest in 32 patients. The mean optical density (OD) of the lungs (800 measurements) was closer to the ideal density with digital radiography (1.45 OD +/- 0.20 [standard deviation] vs 1.75 OD +/- 0.53) and was less often outside the usable range (2.5% vs 42.5%). Receiver operating characteristic analysis for detection of simulated nodules and monitoring devices (nine readers, 4,608 observations) showed that digital radiography was superior to conventional radiography (P less than .05) for four of the nine readers and equivalent to conventional radiography for five readers. The authors concluded that digital radiography produces more consistent and ideal image density and performs at least as well as conventional radiography under phantom test conditions.

Key concepts: Radiography, Medicine, Digital radiography, Imaging phantom, Nuclear medicine, Computed radiography, Standard deviation, Radiology

Related papers

Back to paper searchBrowse research topicsOriginal source
Improved control of image optical density with low-dose digital and conventional radiography in bedside imaging. — Research Paper | ScholarLens