[Evaluation of capability of cell count and detection of tumor cells in cerebrospinal and body fluids by automated hematology analyzer].

Hiroyuki Takemura, Yoko Tabe, Kiyoshi Ishii, Yuki Kobayashi, Y Kuno, Takashi Horii, Kazunori Miyake, Takashi Miida, Akimichi Ohsaka

Open publisher page 7 citations

Abstract

Sysmex XE-5000 offers the body fluid modus which provides the opportunity to count and differentiate leukocytes in body fluids and cerebrospinal fluid (CFS). In this study, we evaluated the basic performance of this application using routinely obtained samples in comparison with manual counting. Reproducibility study yielded good results in samples with a high white blood cell (WBC) count, whereas relatively high imprecision was observed at low WBC counts. Linearity was established up to 1,500 cells/microL in CFS and 5,600 cells/microL in body fluid. The cell count by XE-5000 was highly correlated with that of the microscopic reference method. Highly fluorescent body fluid cells percent (HF-BF%) was observed in samples with tumor cells or activated macrophages, which provides information about the possible presence of tumor cells. In conclusion, total and differential WBC counts in body fluid and CFS can be reliably determined by XE-5000 in samples with increased cell counts. XE-5000 also provides screening information about the presence of tumor cells for further manual examination.

About this research paper

What this paper is about

Sysmex XE-5000 offers the body fluid modus which provides the opportunity to count and differentiate leukocytes in body fluids and cerebrospinal fluid (CFS). In this study, we evaluated the basic performance of this application using routinely obtained samples in comparison with manual counting. Reproducibility study yielded good results in samples with a high white blood cell (WBC) count, whereas relatively high imprecision was observed at low WBC counts. Linearity was established up to 1,500 cells/microL in CFS and 5,600 cells/microL in body fluid. The cell count by XE-5000 was highly correlated with that of the microscopic reference method. Highly fluorescent body fluid cells percent (HF-BF%) was observed in samples with tumor cells or activated macrophages, which provides information about the possible presence of tumor cells. In conclusion, total and differential WBC counts in body fluid and CFS can be reliably determined by XE-5000 in samples with increased cell counts. XE-5000 also provides screening information about the presence of tumor cells for further manual examination.

Why it matters

OpenAlex reports 7 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Sysmex XE-5000 offers the body fluid modus which provides the opportunity to count and differentiate leukocytes in body fluids and cerebrospinal fluid (CFS). In this study, we evaluated the basic performance of this application using routinely obtained samples in comparison with manual counting. Reproducibility study yielded good results in samples with a high white blood cell (WBC) count, whereas relatively high imprecision was observed at low WBC counts. Linearity was established up to 1,500 cells/microL in CFS and 5,600 cells/microL in body fluid. The cell count by XE-5000 was highly correlated with that of the microscopic reference method. Highly fluorescent body fluid cells percent (HF-BF%) was observed in samples with tumor cells or activated macrophages, which provides information about the possible presence of tumor cells. In conclusion, total and differential WBC counts in body fluid and CFS can be reliably determined by XE-5000 in samples with increased cell counts. XE-5000 also provides screening information about the presence of tumor cells for further manual examination.

Key concepts: Hematology analyzer, Body fluid, Cerebrospinal fluid, White blood cell, Cell counting, Pathology, Medicine, Blood cell