Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study
Marcelo D. Campos, Lucio H. Palazzi, Stephan H. Böhm, Gerardo Tusman
Abstract
Open-access reader
Marcelo D. Campos, Lucio H. Palazzi, Stephan H. Böhm, Gerardo Tusman
Abstract
Open-access reader
Abstract Background Volumetric capnography in healthy ventilated neonates showed deformed waveforms, which are supposedly due to technological limitations of flow and carbon dioxide sensors. Aims This bench study analyzed the role of apparatus dead space on the shape of capnograms in simulated neonates with healthy lungs. Methods We simulated mechanical breaths in neonates of 2, 2.5, and 3 kg of body weight using a neonatal volumetric capnography simulator. The simulator was fed by a fixed amount of carbon dioxide of 6 mL/kg/min. Such simulator was ventilated in a volume control mode using fixed ventilatory settings with a tidal volume of 8 mL/kg and respiratory rates of 40, 35, and 30 breaths per minute for the 2, 2.5 and 3 kg neonates, respectively. We tested the above baseline ventilation with and without an additional apparatus dead space of 4 mL. Results Simulations showed that adding the apparatus dead space to baseline ventilation increased the amount of re‐inhaled carbon dioxide in all neonates: 0.16 ± 0.01 to 0.32 ± 0.03 mL (2 kg), 0.14 ± 0.02 to 0.39 ± 0.05 mL (2.5 kg), and 0.13 ± 0.01 to 0.36 ± 0.05 mL (3 kg); (p < .001). Apparatus dead space was computed as part of the airway dead space, and therefore, the ratio of airway dead space to tidal volume increased from 0.51 ± 0.04 to 0.68 ± 0.06, from 0.43 ± 0.04 to 0.62 ± 0.01 and from 0.38 ± 0.01 to 0.60 ± 0.02 in the 2, 2.5 and 3 kg simulated neonates, respectively (p < .001). Compared to baseline ventilation, adding apparatus dead space decreased the ratio of the volume of phase III to VT size from 31% to 11% (2 kg), from 40% to 16% (2.5 kg) and from 50% to 18% (3 kg); (p < .001). Conclusions The addition of a small apparatus dead space artificially deformed the volumetric capnograms in simulated neonates with healthy lungs.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Abstract Background Volumetric capnography in healthy ventilated neonates showed deformed waveforms, which are supposedly due to technological limitations of flow and carbon dioxide sensors. Aims This bench study analyzed the role of apparatus dead space on the shape of capnograms in simulated neonates with healthy lungs. Methods We simulated mechanical breaths in neonates of 2, 2.5, and 3 kg of body weight using a neonatal volumetric capnography simulator. The simulator was fed by a fixed amount of carbon dioxide of 6 mL/kg/min. Such simulator was ventilated in a volume control mode using fixed ventilatory settings with a tidal volume of 8 mL/kg and respiratory rates of 40, 35, and 30 breaths per minute for the 2, 2.5 and 3 kg neonates, respectively. We tested the above baseline ventilation with and without an additional apparatus dead space of 4 mL. Results Simulations showed that adding the apparatus dead space to baseline ventilation increased the amount of re‐inhaled carbon dioxide in all neonates: 0.16 ± 0.01 to 0.32 ± 0.03 mL (2 kg), 0.14 ± 0.02 to 0.39 ± 0.05 mL (2.5 kg), and 0.13 ± 0.01 to 0.36 ± 0.05 mL (3 kg); (p < .001). Apparatus dead space was computed as part of the airway dead space, and therefore, the ratio of airway dead space to tidal volume increased from 0.51 ± 0.04 to 0.68 ± 0.06, from 0.43 ± 0.04 to 0.62 ± 0.01 and from 0.38 ± 0.01 to 0.60 ± 0.02 in the 2, 2.5 and 3 kg simulated neonates, respectively (p < .001). Compared to baseline ventilation, adding apparatus dead space decreased the ratio of the volume of phase III to VT size from 31% to 11% (2 kg), from 40% to 16% (2.5 kg) and from 50% to 18% (3 kg); (p < .001). Conclusions The addition of a small apparatus dead space artificially deformed the volumetric capnograms in simulated neonates with healthy lungs.
Key concepts: Dead space, Medicine, Capnography, Tidal volume, Respiratory minute volume, Ventilation (architecture), Anesthesia, Respiratory rate