2023PLoS ONEOpen access

Reducing contrast-agent volume and radiation dose in CT with 90kVp tube voltage, high tube current modulation, and advanced iteration algorithm

Min Su Park, Hong Il Ha, Jhii-Hyun Ahn, In Jae Lee, Hyun Lim

Open full text 8 citations

Abstract

Increasing utilization of computed tomography (CT) has raised concerns regarding CT radiation dose and technology has been developed to achieve an appropriate balance between image quality, radiation dose, and the amount of contrast material. This study was planned to evaluate the image quality and radiation dose in pancreatic dynamic computed tomography (PDCT) with 90-kVp tube voltage and reduction of the standard amount of contrast agent, compared with 100-kVp PDCT of the research hospital's convention. Total of 51 patients with both CT protocols were included. The average Hounsfield units (HU) values of the abdominal organs and image noise were measured for objective image quality analysis. Two radiologists evaluated five categories of image qualities such as subjective image noise, visibility of small structure, beam hardening or streak artifact, lesion conspicuity and overall diagnostic performance for subjective image quality analysis. The total amount of contrast agent, radiation dose, and image noise decreased in the low-kVp group, by 24.4%, 31.7%, and 20.6%, respectively (p < 0.001). The intraobserver and interobserver agreements were moderate to substantial (k = 0.4-0.8). The contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and figure of merit of the almost organs except psoas muscle in the low-kVp group were significantly higher (p < 0.001). Except for lesion conspicuity, both reviewers judged that subjective image quality of the 90-kVp group was better (p < 0.001). With 90-kVp tube voltage, 25% reduced contrast agent volume with advanced iteration algorithm and high tube current modulation achieved radiation dose reduction of 31.7%, as well as better image quality and diagnostic confidence.

Open-access reader

About this research paper

What this paper is about

Increasing utilization of computed tomography (CT) has raised concerns regarding CT radiation dose and technology has been developed to achieve an appropriate balance between image quality, radiation dose, and the amount of contrast material. This study was planned to evaluate the image quality and radiation dose in pancreatic dynamic computed tomography (PDCT) with 90-kVp tube voltage and reduction of the standard amount of contrast agent, compared with 100-kVp PDCT of the research hospital's convention. Total of 51 patients with both CT protocols were included. The average Hounsfield units (HU) values of the abdominal organs and image noise were measured for objective image quality analysis. Two radiologists evaluated five categories of image qualities such as subjective image noise, visibility of small structure, beam hardening or streak artifact, lesion conspicuity and overall diagnostic performance for subjective image quality analysis. The total amount of contrast agent, radiation dose, and image noise decreased in the low-kVp group, by 24.4%, 31.7%, and 20.6%, respectively (p < 0.001). The intraobserver and interobserver agreements were moderate to substantial (k = 0.4-0.8). The contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and figure of merit of the almost organs except psoas muscle in the low-kVp group were significantly higher (p < 0.001). Except for lesion conspicuity, both reviewers judged that subjective image quality of the 90-kVp group was better (p < 0.001). With 90-kVp tube voltage, 25% reduced contrast agent volume with advanced iteration algorithm and high tube current modulation achieved radiation dose reduction of 31.7%, as well as better image quality and diagnostic confidence.

Why it matters

OpenAlex reports 8 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Increasing utilization of computed tomography (CT) has raised concerns regarding CT radiation dose and technology has been developed to achieve an appropriate balance between image quality, radiation dose, and the amount of contrast material. This study was planned to evaluate the image quality and radiation dose in pancreatic dynamic computed tomography (PDCT) with 90-kVp tube voltage and reduction of the standard amount of contrast agent, compared with 100-kVp PDCT of the research hospital's convention. Total of 51 patients with both CT protocols were included. The average Hounsfield units (HU) values of the abdominal organs and image noise were measured for objective image quality analysis. Two radiologists evaluated five categories of image qualities such as subjective image noise, visibility of small structure, beam hardening or streak artifact, lesion conspicuity and overall diagnostic performance for subjective image quality analysis. The total amount of contrast agent, radiation dose, and image noise decreased in the low-kVp group, by 24.4%, 31.7%, and 20.6%, respectively (p < 0.001). The intraobserver and interobserver agreements were moderate to substantial (k = 0.4-0.8). The contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and figure of merit of the almost organs except psoas muscle in the low-kVp group were significantly higher (p < 0.001). Except for lesion conspicuity, both reviewers judged that subjective image quality of the 90-kVp group was better (p < 0.001). With 90-kVp tube voltage, 25% reduced contrast agent volume with advanced iteration algorithm and high tube current modulation achieved radiation dose reduction of 31.7%, as well as better image quality and diagnostic confidence.

Key concepts: Image quality, Hounsfield scale, Image noise, Contrast-to-noise ratio, Medicine, Nuclear medicine, Contrast (vision), Radiology