Radiation dose and image quality in abdominal CT: Phantom Study
M’hamed El Mansouri, Abdelmajid Choukri, Mohammed Talbi, Othmane KHALLOUKI
Abstract
M’hamed El Mansouri, Abdelmajid Choukri, Mohammed Talbi, Othmane KHALLOUKI
Abstract
The purpose of this study was to evaluate and compare the effect of acquisition parameters of the abdominal computed tomography protocol on image quality and dose quantities, including contrast-to-noise ratio (CNR), noise, spatial resolution (SR) as well as volumetric computed tomography dose index (CTDIv). The measurements have been done on a HITACHI 16-slice scanner using two phantoms (PMMA and Catphan 500), two tube voltages of 120KVp and 100KVp have been analyzed while varying the tube current settings from 155 to 300 (mAs) in order to measure the abdominal image quality quantities and dose (CTDIv). The scanning at 100KVp compared to 120KVp has resulted in a dose reduction up to 38% at mAs ranging from 100 to 300, a decrease in CNR of 28%, a slight decrease in spatial resolution accompanied with an increase in noise. This study has shown that despite the increase in noise at low tube voltages, it is possible to reduce the dose without affecting the quantities of spatial resolution and the contrast/noise ratio by lowering the tube voltage from 120KVp to 100KVp. Keywords: Abdominal CT; CTDIv; Image Quality; Phantom.
OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The purpose of this study was to evaluate and compare the effect of acquisition parameters of the abdominal computed tomography protocol on image quality and dose quantities, including contrast-to-noise ratio (CNR), noise, spatial resolution (SR) as well as volumetric computed tomography dose index (CTDIv). The measurements have been done on a HITACHI 16-slice scanner using two phantoms (PMMA and Catphan 500), two tube voltages of 120KVp and 100KVp have been analyzed while varying the tube current settings from 155 to 300 (mAs) in order to measure the abdominal image quality quantities and dose (CTDIv). The scanning at 100KVp compared to 120KVp has resulted in a dose reduction up to 38% at mAs ranging from 100 to 300, a decrease in CNR of 28%, a slight decrease in spatial resolution accompanied with an increase in noise. This study has shown that despite the increase in noise at low tube voltages, it is possible to reduce the dose without affecting the quantities of spatial resolution and the contrast/noise ratio by lowering the tube voltage from 120KVp to 100KVp. Keywords: Abdominal CT; CTDIv; Image Quality; Phantom.
Key concepts: Imaging phantom, Image quality, Scanner, Image noise, Image resolution, Contrast-to-noise ratio, Nuclear medicine, Noise (video)