2022•Revista gaúcha de enfermagemOpen access

Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets

Natalie Marino Vieira, Maria Paula de Oliveira Pires, Gabriela Beltran Crespo, Larissa Perez Pardo Nascimento, Maria Angélica Sorgini Peterlini, Mavilde da Luz Gonçalves Pedreira

Open full text 2 citations

Abstract

OBJECTIVE: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. METHOD: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump's electronic bolus). Data were analyzed according to mean, standard deviation, Student's t and ANOVA tests (p<0.05). RESULTS: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump's accuracy in all studied rates and manual IV sets filling was similar. CONCLUSION: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.

Open-access reader

About this research paper

What this paper is about

OBJECTIVE: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. METHOD: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump's electronic bolus). Data were analyzed according to mean, standard deviation, Student's t and ANOVA tests (p<0.05). RESULTS: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump's accuracy in all studied rates and manual IV sets filling was similar. CONCLUSION: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.

Why it matters

OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. METHOD: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump's electronic bolus). Data were analyzed according to mean, standard deviation, Student's t and ANOVA tests (p<0.05). RESULTS: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump's accuracy in all studied rates and manual IV sets filling was similar. CONCLUSION: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.

Key concepts: Syringe, Priming (agriculture), Syringe driver, Anesthesia, Medicine, Biology, Botany, Germination

Related papers

Back to paper searchBrowse research topicsOriginal source
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets — Research Paper | ScholarLens