Start‐up delays of infusion syringe pumps
Tobias Neff, Joachim E. Fischer, S Fehr, O. Baenziger, M Weiss
Abstract
Tobias Neff, Joachim E. Fischer, S Fehr, O. Baenziger, M Weiss
Abstract
BACKGROUND: We performed a bench experiment to investigate the extent of start-up delays in fluid delivery for four different syringe pumps after initially placing the infusion syringe in the syringe pump. METHODS: Pump performance was determined at an infusion rate of 1 ml.h-1 with and without a fluid bolus delivered by the infusion pump prior to connecting the infusion line to the simulated patient. RESULTS: The time (mean +/- SD) from starting the pump up to first fluid delivery (t1) differed considerably between pumps (from 6.75 +/- 4.4 to 57.2 +/- 28.6 min) as did the time to steady state fluid delivery (t2) (from 19.6 +/- 9.3 to 76.3 +/- 29.0 min). Applying an initial bolus of 2 ml before connecting the line to the simulated patient practically eliminated the delay in fluid delivery (t1 ranging from 0.3 +/- 0.1 to 1.1 +/- 0.8 min). This manoeuvre also reduced the time to steady flow delivery (t2 from 6.0 +/- 3.1 to 11.1 +/- 4.3 min, P<0.001) and minimized the differences between syringe pumps. CONCLUSIONS: Syringe pump design affects start-up delay times because of free play of the syringe. These delays can be eliminated by a start-up bolus of 2 ml prior to connecting the infusion line to the patient.
OpenAlex reports 86 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
BACKGROUND: We performed a bench experiment to investigate the extent of start-up delays in fluid delivery for four different syringe pumps after initially placing the infusion syringe in the syringe pump. METHODS: Pump performance was determined at an infusion rate of 1 ml.h-1 with and without a fluid bolus delivered by the infusion pump prior to connecting the infusion line to the simulated patient. RESULTS: The time (mean +/- SD) from starting the pump up to first fluid delivery (t1) differed considerably between pumps (from 6.75 +/- 4.4 to 57.2 +/- 28.6 min) as did the time to steady state fluid delivery (t2) (from 19.6 +/- 9.3 to 76.3 +/- 29.0 min). Applying an initial bolus of 2 ml before connecting the line to the simulated patient practically eliminated the delay in fluid delivery (t1 ranging from 0.3 +/- 0.1 to 1.1 +/- 0.8 min). This manoeuvre also reduced the time to steady flow delivery (t2 from 6.0 +/- 3.1 to 11.1 +/- 4.3 min, P<0.001) and minimized the differences between syringe pumps. CONCLUSIONS: Syringe pump design affects start-up delay times because of free play of the syringe. These delays can be eliminated by a start-up bolus of 2 ml prior to connecting the infusion line to the patient.
Key concepts: Syringe, Syringe driver, Infusion pump, Medicine, Bolus (digestion), Anesthesia, Continuous infusion, Peristaltic pump