Application value of low radiation dose coronary angiography using 320-slice volume CT

Yuning Pan, Qiuli Huang, Dawei Ren, An‐Le Wu, Xianwang Ye, Jie Zhang

Open publisher page 0 citations

Abstract

Objective To evaluate the image quality and radiation dose of prospective electrocardiography-triggered coronary 320-slice volume CT angiography with different kV, and the feasibility of coronary scan with < 1 mSv radiation dose.Methods Eighty consecutive patients were randomly divided into two groups equally.The tube voltage according to paradigm was 100 kV in group A and 120 kV in group B.All raw data in group A was reconstructed by the software AIDR in CT system to create a new group named as A1. Such parameters as the mean intraluminal attenuation (SI),noise (SD),signal-to-noise ratio (SNR),contrast-to-noise ratio (CNR),effective radiation dose(E) and image quality score measured in group A were compared with those in group B.The values such as SI,SD,SNR,CNR,image quality scores were compared between group A and group A1.The significance of group B and group A1 was compared in SI,SD,SNR,CNR,image quality scores as well.Results E in group A was significantly lower than that in group B[ E =(0.67 ± 0.18) mSv in group A vs.E =(3.08 ± 1.04) mSv in group B].The value of E in group A was decreased by 78% compared to group B(t =- 14.30,P<0.05 ).There was no significant difference in mean image quality scores between two groups(4.57 ± 0.57in groupA vs.4.59 ± 0.59 in group B,t=-1.17,P>0.05).The values of SI,SD,SNR,CNR in group A were (570.8 ±131.5)HU,25.1 ±6.9,24.5 ±9.1,19.8 ±6.1.And the values of SI,SD,SNR,CNR in group B were (460.6 ± 14.3) HU,15.1 ±3.6,31.7 ±7.7,29.3 ±6.8.The values of SI and SD in group A were significantly higher than those in group B(t =4.49,8.18,P <0.05). The values of SNR and CNR in group A were lower than those in group B (t =-4.24,-6.19,P<0.05).The valuesofS1,SD,SNR,CNR,image quality scores in group Al were (557.9 ±24.5) HU,21.1 ±6.0,27.7±10.0,23.4±7.8,4.60 ± 0.56.There was no difference in the SI and the image quality scores between group A and group A1 ( t =1.09,- 1.90,P > 0.05).Conclusions 320-slice volume CT with 100 kV tube voltage and prospective ECG-triggered technique can reduce the radiation dose to less than 1 mSv and obtain optimal images in diagnosis of coronary arterial diseases. Key words: 320-slice volume CT;  Coronary angiography;  Noise;  Radiation dose

About this research paper

What this paper is about

Objective To evaluate the image quality and radiation dose of prospective electrocardiography-triggered coronary 320-slice volume CT angiography with different kV, and the feasibility of coronary scan with < 1 mSv radiation dose.Methods Eighty consecutive patients were randomly divided into two groups equally.The tube voltage according to paradigm was 100 kV in group A and 120 kV in group B.All raw data in group A was reconstructed by the software AIDR in CT system to create a new group named as A1. Such parameters as the mean intraluminal attenuation (SI),noise (SD),signal-to-noise ratio (SNR),contrast-to-noise ratio (CNR),effective radiation dose(E) and image quality score measured in group A were compared with those in group B.The values such as SI,SD,SNR,CNR,image quality scores were compared between group A and group A1.The significance of group B and group A1 was compared in SI,SD,SNR,CNR,image quality scores as well.Results E in group A was significantly lower than that in group B[ E =(0.67 ± 0.18) mSv in group A vs.E =(3.08 ± 1.04) mSv in group B].The value of E in group A was decreased by 78% compared to group B(t =- 14.30,P<0.05 ).There was no significant difference in mean image quality scores between two groups(4.57 ± 0.57in groupA vs.4.59 ± 0.59 in group B,t=-1.17,P>0.05).The values of SI,SD,SNR,CNR in group A were (570.8 ±131.5)HU,25.1 ±6.9,24.5 ±9.1,19.8 ±6.1.And the values of SI,SD,SNR,CNR in group B were (460.6 ± 14.3) HU,15.1 ±3.6,31.7 ±7.7,29.3 ±6.8.The values of SI and SD in group A were significantly higher than those in group B(t =4.49,8.18,P <0.05). The values of SNR and CNR in group A were lower than those in group B (t =-4.24,-6.19,P<0.05).The valuesofS1,SD,SNR,CNR,image quality scores in group Al were (557.9 ±24.5) HU,21.1 ±6.0,27.7±10.0,23.4±7.8,4.60 ± 0.56.There was no difference in the SI and the image quality scores between group A and group A1 ( t =1.09,- 1.90,P > 0.05).Conclusions 320-slice volume CT with 100 kV tube voltage and prospective ECG-triggered technique can reduce the radiation dose to less than 1 mSv and obtain optimal images in diagnosis of coronary arterial diseases. Key words: 320-slice volume CT;  Coronary angiography;  Noise;  Radiation dose

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the image quality and radiation dose of prospective electrocardiography-triggered coronary 320-slice volume CT angiography with different kV, and the feasibility of coronary scan with < 1 mSv radiation dose.Methods Eighty consecutive patients were randomly divided into two groups equally.The tube voltage according to paradigm was 100 kV in group A and 120 kV in group B.All raw data in group A was reconstructed by the software AIDR in CT system to create a new group named as A1. Such parameters as the mean intraluminal attenuation (SI),noise (SD),signal-to-noise ratio (SNR),contrast-to-noise ratio (CNR),effective radiation dose(E) and image quality score measured in group A were compared with those in group B.The values such as SI,SD,SNR,CNR,image quality scores were compared between group A and group A1.The significance of group B and group A1 was compared in SI,SD,SNR,CNR,image quality scores as well.Results E in group A was significantly lower than that in group B[ E =(0.67 ± 0.18) mSv in group A vs.E =(3.08 ± 1.04) mSv in group B].The value of E in group A was decreased by 78% compared to group B(t =- 14.30,P<0.05 ).There was no significant difference in mean image quality scores between two groups(4.57 ± 0.57in groupA vs.4.59 ± 0.59 in group B,t=-1.17,P>0.05).The values of SI,SD,SNR,CNR in group A were (570.8 ±131.5)HU,25.1 ±6.9,24.5 ±9.1,19.8 ±6.1.And the values of SI,SD,SNR,CNR in group B were (460.6 ± 14.3) HU,15.1 ±3.6,31.7 ±7.7,29.3 ±6.8.The values of SI and SD in group A were significantly higher than those in group B(t =4.49,8.18,P <0.05). The values of SNR and CNR in group A were lower than those in group B (t =-4.24,-6.19,P<0.05).The valuesofS1,SD,SNR,CNR,image quality scores in group Al were (557.9 ±24.5) HU,21.1 ±6.0,27.7±10.0,23.4±7.8,4.60 ± 0.56.There was no difference in the SI and the image quality scores between group A and group A1 ( t =1.09,- 1.90,P > 0.05).Conclusions 320-slice volume CT with 100 kV tube voltage and prospective ECG-triggered technique can reduce the radiation dose to less than 1 mSv and obtain optimal images in diagnosis of coronary arterial diseases. Key words: 320-slice volume CT;  Coronary angiography;  Noise;  Radiation dose

Key concepts: Nuclear medicine, Image quality, Medicine, Image noise, Group B, Effective dose (radiation), Contrast-to-noise ratio, Radiation dose

Related papers

Back to paper searchBrowse research topicsOriginal source
Application value of low radiation dose coronary angiography using 320-slice volume CT — Research Paper | ScholarLens