Clinical application of low-dose CT scan technique in orbit
Feng-yin Feng, Zhao-hua Zheng, Jia-xiu Deng
Abstract
Feng-yin Feng, Zhao-hua Zheng, Jia-xiu Deng
Abstract
Objective To explore the effects of low-dose CT scan on diagnostic quality of images and the radiation dose to the lens.Methods 60 cases were scanned in axial orientation at different mA.Lens radiation dose was measured and image quality was evaluated.Results Image quality score of group A was(11.2 ± 0.7)points,higher than the group C(4.9 ± 1.8)points(t =12.634,P <0.05),and score difference of group B(10.6 ± 1.1)was significantly different(P > 0.05).CTDIvol of group A,group B and group C were 18.32mGy,7.63mGy and 4.58mGy,group B,group C and group A difference was statistically significant(t =15.625 and 14.037,P < 0.05).Group B and C groups difference was not statistically significant(P > 0.05).Conclusion Low dose CT scanning(50mA)in orbit not only could assure the image quality,but also decrease the radiation dose. Key words: Tomography,X-ary computed; Radiation dosage; Orbit
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To explore the effects of low-dose CT scan on diagnostic quality of images and the radiation dose to the lens.Methods 60 cases were scanned in axial orientation at different mA.Lens radiation dose was measured and image quality was evaluated.Results Image quality score of group A was(11.2 ± 0.7)points,higher than the group C(4.9 ± 1.8)points(t =12.634,P <0.05),and score difference of group B(10.6 ± 1.1)was significantly different(P > 0.05).CTDIvol of group A,group B and group C were 18.32mGy,7.63mGy and 4.58mGy,group B,group C and group A difference was statistically significant(t =15.625 and 14.037,P < 0.05).Group B and C groups difference was not statistically significant(P > 0.05).Conclusion Low dose CT scanning(50mA)in orbit not only could assure the image quality,but also decrease the radiation dose. Key words: Tomography,X-ary computed; Radiation dosage; Orbit
Key concepts: Medicine, Nuclear medicine, Radiation dose, Image quality, Orbit (dynamics), Computed tomography, Significant difference, Group B