Evaluation of serum gamma-glutamyltransferase assays by using the International Federation of Clinical Chemistry and Laboratory Medicine reference method

Ning Xu, Songbai Zheng, Li Lin, Junhua Zhuang, Jianhua Xu, Lian-ying Lin, Lei Sun, Long-hua Guo, Xianzhang Huang

Open publisher page 0 citations

Abstract

Objective To evaluate the measurement accuracy of serum gamma-glutamyltransferase (GGT) assays manufactured in China. Methods The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for GGT was set up and, after verification, was used to evaluate the performance of routine assay systems made in China. The evaluation was performed twice before and after a calibration by a common serum calibrator. Results For the reference measurement, the within run and total CVs were all less than 1%. The biases with the target values of IFCC External Quality Assessment Scheme for Reference Laboratories (RELA) were all within the limit of equivalence. Before a calibration with a common calibrator, the largest biases of results of GGT of the routine tasting systems compared with reference method at three medical decide levels were -47.53%, -34.11% and -30.07% respectively, and the averaged biases were 14.53% ,12.88% and 12.48%. After calibrating by fresh serum calibrator,the largest biases were reduced to - 17.63%, -5.88% and -4.08% ,the averaged biases were reduced to 7.50%, 2.70% and 1.87%. Conclusion The performance of GGT measurements can be effectively improved by using a common fresh serum calibrator that has a value assigned with the reference method. Key words: gamma-Glutamyltransferase;  Reference values;  Indicators and reagents;  Calibration

About this research paper

What this paper is about

Objective To evaluate the measurement accuracy of serum gamma-glutamyltransferase (GGT) assays manufactured in China. Methods The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for GGT was set up and, after verification, was used to evaluate the performance of routine assay systems made in China. The evaluation was performed twice before and after a calibration by a common serum calibrator. Results For the reference measurement, the within run and total CVs were all less than 1%. The biases with the target values of IFCC External Quality Assessment Scheme for Reference Laboratories (RELA) were all within the limit of equivalence. Before a calibration with a common calibrator, the largest biases of results of GGT of the routine tasting systems compared with reference method at three medical decide levels were -47.53%, -34.11% and -30.07% respectively, and the averaged biases were 14.53% ,12.88% and 12.48%. After calibrating by fresh serum calibrator,the largest biases were reduced to - 17.63%, -5.88% and -4.08% ,the averaged biases were reduced to 7.50%, 2.70% and 1.87%. Conclusion The performance of GGT measurements can be effectively improved by using a common fresh serum calibrator that has a value assigned with the reference method. Key words: gamma-Glutamyltransferase;  Reference values;  Indicators and reagents;  Calibration

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the measurement accuracy of serum gamma-glutamyltransferase (GGT) assays manufactured in China. Methods The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for GGT was set up and, after verification, was used to evaluate the performance of routine assay systems made in China. The evaluation was performed twice before and after a calibration by a common serum calibrator. Results For the reference measurement, the within run and total CVs were all less than 1%. The biases with the target values of IFCC External Quality Assessment Scheme for Reference Laboratories (RELA) were all within the limit of equivalence. Before a calibration with a common calibrator, the largest biases of results of GGT of the routine tasting systems compared with reference method at three medical decide levels were -47.53%, -34.11% and -30.07% respectively, and the averaged biases were 14.53% ,12.88% and 12.48%. After calibrating by fresh serum calibrator,the largest biases were reduced to - 17.63%, -5.88% and -4.08% ,the averaged biases were reduced to 7.50%, 2.70% and 1.87%. Conclusion The performance of GGT measurements can be effectively improved by using a common fresh serum calibrator that has a value assigned with the reference method. Key words: gamma-Glutamyltransferase;  Reference values;  Indicators and reagents;  Calibration

Key concepts: Reference values, Calibration, External quality assessment, Medical laboratory, Mathematics, Nuclear medicine, Statistics, Chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
Evaluation of serum gamma-glutamyltransferase assays by using the International Federation of Clinical Chemistry and Laboratory Medicine reference method — Research Paper | ScholarLens