Performance verification and evaluation of a one-step HBV DNA quantitative detecting system

Wei Ying, Wan‐Chun Li, Tao Song, Jue Wang, Fugen Mo, Yipin Wang

Open publisher page 0 citations

Abstract

Objective To evaluate analytical performance and clinical application value of a one-step HBV DNA quantitative detecting system. Methods Analytical performance of the one-step HBV DNA quantitative detecting reagents included precision, residual contamination, accuracy, functional sensitivity and analytical measurement range were verified by collecting high concentration samples and external quality control samples from Jiangsu provincial clinical test center. Results The within-run coefficient of variation (CV) of both low and high concentration samples were below 5%, meanwhile the intra-assay CV was below 3/5 TEa and inter-assay CV was below 4/5 TEa. There was no residual contamination and the analytic accuracy met the requirement of external quality assessment (EQA). Functional sensitivity was able to attain 100 IU/ml, while the day to day CV was below 20%. It exhibited a benign linear relation from 7.58×101 to 7.58×108 IU/ml. Conclusions The analytic performance of a new testing system must be evaluated particularly before detecting samples of patients by quantitative tests. This study proves that the one-step HBV DNA quantitative detecting reagents can meet requirement of hepatitis B screening and clinical therapy monitoring, which is economic and simple for clinical routine tests. Key words: Hepatitis B virus; Quantitative detecting system; Performance verification; Fluorescence quantitative PCR

About this research paper

What this paper is about

Objective To evaluate analytical performance and clinical application value of a one-step HBV DNA quantitative detecting system. Methods Analytical performance of the one-step HBV DNA quantitative detecting reagents included precision, residual contamination, accuracy, functional sensitivity and analytical measurement range were verified by collecting high concentration samples and external quality control samples from Jiangsu provincial clinical test center. Results The within-run coefficient of variation (CV) of both low and high concentration samples were below 5%, meanwhile the intra-assay CV was below 3/5 TEa and inter-assay CV was below 4/5 TEa. There was no residual contamination and the analytic accuracy met the requirement of external quality assessment (EQA). Functional sensitivity was able to attain 100 IU/ml, while the day to day CV was below 20%. It exhibited a benign linear relation from 7.58×101 to 7.58×108 IU/ml. Conclusions The analytic performance of a new testing system must be evaluated particularly before detecting samples of patients by quantitative tests. This study proves that the one-step HBV DNA quantitative detecting reagents can meet requirement of hepatitis B screening and clinical therapy monitoring, which is economic and simple for clinical routine tests. Key words: Hepatitis B virus; Quantitative detecting system; Performance verification; Fluorescence quantitative PCR

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate analytical performance and clinical application value of a one-step HBV DNA quantitative detecting system. Methods Analytical performance of the one-step HBV DNA quantitative detecting reagents included precision, residual contamination, accuracy, functional sensitivity and analytical measurement range were verified by collecting high concentration samples and external quality control samples from Jiangsu provincial clinical test center. Results The within-run coefficient of variation (CV) of both low and high concentration samples were below 5%, meanwhile the intra-assay CV was below 3/5 TEa and inter-assay CV was below 4/5 TEa. There was no residual contamination and the analytic accuracy met the requirement of external quality assessment (EQA). Functional sensitivity was able to attain 100 IU/ml, while the day to day CV was below 20%. It exhibited a benign linear relation from 7.58×101 to 7.58×108 IU/ml. Conclusions The analytic performance of a new testing system must be evaluated particularly before detecting samples of patients by quantitative tests. This study proves that the one-step HBV DNA quantitative detecting reagents can meet requirement of hepatitis B screening and clinical therapy monitoring, which is economic and simple for clinical routine tests. Key words: Hepatitis B virus; Quantitative detecting system; Performance verification; Fluorescence quantitative PCR

Key concepts: Coefficient of variation, Hepatitis B virus, Residual, Quantitative analysis (chemistry), Contamination, Real-time polymerase chain reaction, External quality assessment, Reproducibility

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance verification and evaluation of a one-step HBV DNA quantitative detecting system — Research Paper | ScholarLens