Investigation of optimum exposure dose for chest imaging using CR and amorphous silocon DR system

Guoquan Cao, Huazhi Xu, Yunpeng Tai, Wu Enfu, Zheng Xiangwu

Open publisher page 0 citations

Abstract

Objective To compare the difference of entrance dose between CR and amorphous silocon DR system in chest imaging, and to discuss their optimum exposure dose. Methods For CR and DR, different entrance dose was measured by dosimeter in chest phantom. The value of IQFinv was analyzed by CDRAD2. 0 software. Image quality difference between CR and DR was assessed by group t-test. The relationship between image quality and entrance dose was tested by using Pearson correlation analysis. The best IQFinv values in CR and DR system were achieved via ROC curve analysis, and the exposure dose was then calculated. Results There were direct correlation values between entrance dose and the value of IQFinv in CR and DR system, respectively( r =0. 893 ,0. 848 ,P < 0. 01 ) . The linear regression equation for DR was IQFinv =0. 0050 +3. 359, and for CR was IQFinv =0. 005D + I. 651 , where D was entrance dose. The difference of IQFinv value between CR and DR was significant(t = 5. 455 ,P < 0. 05). The best IQFinv value of the two groups from ROC analysis was 3.55. Conclusions With the entrance dose increased, the detection ability of contrast-detail was elevated in the two digital radiography systems. With equal entrance dose, the detection ability of DR in contrast-detail was superior to CR. With equal image quality, DR obviously decreased the radiation dose to the patients. Key words: Digital radiography; Image quality; Exposure dose; Quality control

About this research paper

What this paper is about

Objective To compare the difference of entrance dose between CR and amorphous silocon DR system in chest imaging, and to discuss their optimum exposure dose. Methods For CR and DR, different entrance dose was measured by dosimeter in chest phantom. The value of IQFinv was analyzed by CDRAD2. 0 software. Image quality difference between CR and DR was assessed by group t-test. The relationship between image quality and entrance dose was tested by using Pearson correlation analysis. The best IQFinv values in CR and DR system were achieved via ROC curve analysis, and the exposure dose was then calculated. Results There were direct correlation values between entrance dose and the value of IQFinv in CR and DR system, respectively( r =0. 893 ,0. 848 ,P < 0. 01 ) . The linear regression equation for DR was IQFinv =0. 0050 +3. 359, and for CR was IQFinv =0. 005D + I. 651 , where D was entrance dose. The difference of IQFinv value between CR and DR was significant(t = 5. 455 ,P < 0. 05). The best IQFinv value of the two groups from ROC analysis was 3.55. Conclusions With the entrance dose increased, the detection ability of contrast-detail was elevated in the two digital radiography systems. With equal entrance dose, the detection ability of DR in contrast-detail was superior to CR. With equal image quality, DR obviously decreased the radiation dose to the patients. Key words: Digital radiography; Image quality; Exposure dose; Quality control

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To compare the difference of entrance dose between CR and amorphous silocon DR system in chest imaging, and to discuss their optimum exposure dose. Methods For CR and DR, different entrance dose was measured by dosimeter in chest phantom. The value of IQFinv was analyzed by CDRAD2. 0 software. Image quality difference between CR and DR was assessed by group t-test. The relationship between image quality and entrance dose was tested by using Pearson correlation analysis. The best IQFinv values in CR and DR system were achieved via ROC curve analysis, and the exposure dose was then calculated. Results There were direct correlation values between entrance dose and the value of IQFinv in CR and DR system, respectively( r =0. 893 ,0. 848 ,P < 0. 01 ) . The linear regression equation for DR was IQFinv =0. 0050 +3. 359, and for CR was IQFinv =0. 005D + I. 651 , where D was entrance dose. The difference of IQFinv value between CR and DR was significant(t = 5. 455 ,P < 0. 05). The best IQFinv value of the two groups from ROC analysis was 3.55. Conclusions With the entrance dose increased, the detection ability of contrast-detail was elevated in the two digital radiography systems. With equal entrance dose, the detection ability of DR in contrast-detail was superior to CR. With equal image quality, DR obviously decreased the radiation dose to the patients. Key words: Digital radiography; Image quality; Exposure dose; Quality control

Key concepts: Nuclear medicine, Digital radiography, Imaging phantom, Image quality, Dosimeter, Medicine, Radiation dose, Significant difference

Related papers

Back to paper searchBrowse research topicsOriginal source
Investigation of optimum exposure dose for chest imaging using CR and amorphous silocon DR system — Research Paper | ScholarLens