Flexible film dosimeter for in vivo dosimetry

Jin Dong Cho, Jaeman Son, Jiwon Sung, Chang Heon Choi, Jin Sung Kim, Jin Sung Kim, Hong‐Gyun Wu, Jong Min Park, Jung‐in Kim, Jung‐in Kim

Open publisher page 25 citations

Abstract

PURPOSE: The aims of this study were to develop a flexible film dosimeter applicable to the irregular surface of a patient for in vivo dosimetry and to evaluate the device's dosimetric characteristics. METHODS: A flexible film dosimeter with active layers consisting of radiochromic-sensitive films and flexible silicone materials was constructed. The dose-response, sensitivity, scanning orientation dependence, energy dependence, and dose rate dependence of the flexible film dosimeter were tested. Irradiated dosimeters were scanned 24 h post-irradiation, and the region of interest was 5 mm × 5 mm. Biological stability tests ensured the safety of application of the flexible film dosimeter for patients. A preliminary clinical study with the flexible film dosimeter was implemented on four patients. RESULTS: The red channel demonstrated the highest sensitivity among all channels, and the response sensitivity of the dosimeter decreased with the applied dose, which were the same as the characteristics of GAFCHROMIC EBT3 radiochromic films. The flexible film dosimeter showed no significant energy dependence for photon beams of 6 MV, 6 MV flattening filter-free (FFF), 10 MV, and 15 MV. The flexible film dosimeter showed no substantial dose rate dependence with 6 or 6 MV FFF. In terms of biological stability, the flexible film dosimeter demonstrated no cytotoxicity, no irritation, and no skin sensitization. In the preliminary clinical study, the dose differences between the measurements with the flexible film dosimeter and calculations with the treatment planning system ranged from -0.1% to 1.2% for all patients. CONCLUSIONS: The dosimeter developed in this study is a flexible film capable of attachment to a curved skin surface. The biological test results indicate the stability of the flexible film dosimeter. The preliminary clinical study showed that the flexible film dosimeter can be successfully applied as an in vivo dosimeter.

About this research paper

What this paper is about

PURPOSE: The aims of this study were to develop a flexible film dosimeter applicable to the irregular surface of a patient for in vivo dosimetry and to evaluate the device's dosimetric characteristics. METHODS: A flexible film dosimeter with active layers consisting of radiochromic-sensitive films and flexible silicone materials was constructed. The dose-response, sensitivity, scanning orientation dependence, energy dependence, and dose rate dependence of the flexible film dosimeter were tested. Irradiated dosimeters were scanned 24 h post-irradiation, and the region of interest was 5 mm × 5 mm. Biological stability tests ensured the safety of application of the flexible film dosimeter for patients. A preliminary clinical study with the flexible film dosimeter was implemented on four patients. RESULTS: The red channel demonstrated the highest sensitivity among all channels, and the response sensitivity of the dosimeter decreased with the applied dose, which were the same as the characteristics of GAFCHROMIC EBT3 radiochromic films. The flexible film dosimeter showed no significant energy dependence for photon beams of 6 MV, 6 MV flattening filter-free (FFF), 10 MV, and 15 MV. The flexible film dosimeter showed no substantial dose rate dependence with 6 or 6 MV FFF. In terms of biological stability, the flexible film dosimeter demonstrated no cytotoxicity, no irritation, and no skin sensitization. In the preliminary clinical study, the dose differences between the measurements with the flexible film dosimeter and calculations with the treatment planning system ranged from -0.1% to 1.2% for all patients. CONCLUSIONS: The dosimeter developed in this study is a flexible film capable of attachment to a curved skin surface. The biological test results indicate the stability of the flexible film dosimeter. The preliminary clinical study showed that the flexible film dosimeter can be successfully applied as an in vivo dosimeter.

Why it matters

OpenAlex reports 25 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: The aims of this study were to develop a flexible film dosimeter applicable to the irregular surface of a patient for in vivo dosimetry and to evaluate the device's dosimetric characteristics. METHODS: A flexible film dosimeter with active layers consisting of radiochromic-sensitive films and flexible silicone materials was constructed. The dose-response, sensitivity, scanning orientation dependence, energy dependence, and dose rate dependence of the flexible film dosimeter were tested. Irradiated dosimeters were scanned 24 h post-irradiation, and the region of interest was 5 mm × 5 mm. Biological stability tests ensured the safety of application of the flexible film dosimeter for patients. A preliminary clinical study with the flexible film dosimeter was implemented on four patients. RESULTS: The red channel demonstrated the highest sensitivity among all channels, and the response sensitivity of the dosimeter decreased with the applied dose, which were the same as the characteristics of GAFCHROMIC EBT3 radiochromic films. The flexible film dosimeter showed no significant energy dependence for photon beams of 6 MV, 6 MV flattening filter-free (FFF), 10 MV, and 15 MV. The flexible film dosimeter showed no substantial dose rate dependence with 6 or 6 MV FFF. In terms of biological stability, the flexible film dosimeter demonstrated no cytotoxicity, no irritation, and no skin sensitization. In the preliminary clinical study, the dose differences between the measurements with the flexible film dosimeter and calculations with the treatment planning system ranged from -0.1% to 1.2% for all patients. CONCLUSIONS: The dosimeter developed in this study is a flexible film capable of attachment to a curved skin surface. The biological test results indicate the stability of the flexible film dosimeter. The preliminary clinical study showed that the flexible film dosimeter can be successfully applied as an in vivo dosimeter.

Key concepts: Dosimeter, Dosimetry, Materials science, Irradiation, Sensitivity (control systems), Biomedical engineering, Optoelectronics, Optics