Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group

Min Chen, Lina Zhang, Jia Xu, Xinyi Chen, Yuxiang Gu, Yuping Ren, Kaijun Wang

Open full text 84 citations

Abstract

BACKGROUND: Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. METHODS: This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. RESULTS: The mean difference (Δ) of NCT-GAT did not differ from (Δ) iCare-GAT in IOP < 10 and 10-21 mmHg group. However, (Δ) NCT-GAT was significantly higher than (Δ) iCare-GAT in IOP 22-30 and > 30 mmHg group (P < 0.05). Bland-Altman analysis showed significant agreement between the three devices (P < 0.01). IOP measurements of the three methods were significantly correlated with CCT (P < 0.01). CONCLUSIONS: ICare pro shows a higher agreement with GAT over a wide range of IOP compared with NCT. The consistency between the three tonometers was similar in a low and normal IOP range. However, NCT shows a greater overestimate of IOP in moderate and higher IOP group. The variability of IOP measurement affected by CCT is NCT > iCare pro > GAT.

Open-access reader

About this research paper

What this paper is about

BACKGROUND: Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. METHODS: This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. RESULTS: The mean difference (Δ) of NCT-GAT did not differ from (Δ) iCare-GAT in IOP < 10 and 10-21 mmHg group. However, (Δ) NCT-GAT was significantly higher than (Δ) iCare-GAT in IOP 22-30 and > 30 mmHg group (P < 0.05). Bland-Altman analysis showed significant agreement between the three devices (P < 0.01). IOP measurements of the three methods were significantly correlated with CCT (P < 0.01). CONCLUSIONS: ICare pro shows a higher agreement with GAT over a wide range of IOP compared with NCT. The consistency between the three tonometers was similar in a low and normal IOP range. However, NCT shows a greater overestimate of IOP in moderate and higher IOP group. The variability of IOP measurement affected by CCT is NCT > iCare pro > GAT.

Why it matters

OpenAlex reports 84 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. METHODS: This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. RESULTS: The mean difference (Δ) of NCT-GAT did not differ from (Δ) iCare-GAT in IOP < 10 and 10-21 mmHg group. However, (Δ) NCT-GAT was significantly higher than (Δ) iCare-GAT in IOP 22-30 and > 30 mmHg group (P < 0.05). Bland-Altman analysis showed significant agreement between the three devices (P < 0.01). IOP measurements of the three methods were significantly correlated with CCT (P < 0.01). CONCLUSIONS: ICare pro shows a higher agreement with GAT over a wide range of IOP compared with NCT. The consistency between the three tonometers was similar in a low and normal IOP range. However, NCT shows a greater overestimate of IOP in moderate and higher IOP group. The variability of IOP measurement affected by CCT is NCT > iCare pro > GAT.

Key concepts: Medicine, Goldmann Applanation Tonometer, Intraocular pressure, Ophthalmology, Intraclass correlation, Glaucoma, Optometry, Psychometrics