Pseudoaneurysm was caused at the puncture site of the left groin after percutaneous coronary intervention. Balloon tamponade was attempted for hemostasis at the aneurysmal site. However, hemostasis was not achieved. Next, direct puncture of the pseudoaneurysm was tried. A 0.014-inch guidewire was crossed from the neck of the pseudoaneurysm to the left common femoral artery. The wire was replaced with a 0.035-inch guidewire. An 8F Angio-Seal (Terumo Interventional Systems, Somerset, NJ) was inserted, and a collagen plug was deployed at the neck of the pseudoaneurysm. Final angiography revealed completion of hemostasis. Three-dimensional computed tomography angiography after 8 months revealed no evidence of recurrence.

About this research paper

What this paper is about

Pseudoaneurysm was caused at the puncture site of the left groin after percutaneous coronary intervention. Balloon tamponade was attempted for hemostasis at the aneurysmal site. However, hemostasis was not achieved. Next, direct puncture of the pseudoaneurysm was tried. A 0.014-inch guidewire was crossed from the neck of the pseudoaneurysm to the left common femoral artery. The wire was replaced with a 0.035-inch guidewire. An 8F Angio-Seal (Terumo Interventional Systems, Somerset, NJ) was inserted, and a collagen plug was deployed at the neck of the pseudoaneurysm. Final angiography revealed completion of hemostasis. Three-dimensional computed tomography angiography after 8 months revealed no evidence of recurrence.

Why it matters

OpenAlex reports 4 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Pseudoaneurysm was caused at the puncture site of the left groin after percutaneous coronary intervention. Balloon tamponade was attempted for hemostasis at the aneurysmal site. However, hemostasis was not achieved. Next, direct puncture of the pseudoaneurysm was tried. A 0.014-inch guidewire was crossed from the neck of the pseudoaneurysm to the left common femoral artery. The wire was replaced with a 0.035-inch guidewire. An 8F Angio-Seal (Terumo Interventional Systems, Somerset, NJ) was inserted, and a collagen plug was deployed at the neck of the pseudoaneurysm. Final angiography revealed completion of hemostasis. Three-dimensional computed tomography angiography after 8 months revealed no evidence of recurrence.

Key concepts: Pseudoaneurysm, Medicine, Hemostasis, Vascular closure device, Femoral artery, Surgery, Percutaneous coronary intervention, Groin

Related papers

Back to paper searchBrowse research topicsOriginal source
Off-label use of Angio-Seal vascular closure device for the repair of femoral pseudoaneurysm after transfemoral coronary intervention — Research Paper | ScholarLens