Measured X-ray Spectra and Absorbed Dose Conversion Factors in a Water Equivalent Phantom Irradiated with Diagnostic X-rays

Kiyoshi Aoki

Open full text 0 citations

Abstract

The diagnostic X-ray spectra in a water equivalent phantom have been measured. From these measured spectra, the absorbed dose conversion factors of water were derived. The primary X-ray spectra were also measured and the scattered X-ray spectra were calculated by subtraction. The measurements were made at the depths of 0, 5, 10, 15, and 20 cm in a 20 cm-thick phantom and at the X-ray tube voltages 60, 90, and 120 kV by using a small silicon diode detector. The radiation field size was 30×30 cm2 at the phantom surface. In the obtained spectra, the fraction of the scattered photon number is increased with the depth. The X-ray qualities of the spectra in the phantom were near the qualities of primary X-rays when the depth is 0 cm, and became near the qualities of scattered X-rays as the depth increases. The changes of the X-ray qualities due to the depth change were small; photon mean-energy changes were within 4.6 keV. The changes in the absorbed dose conversion factors were also small (within 0.68%). These conversion factors were 0.4–2.3% larger than those obtained from the effective energy of incident X-rays and only −0.3 to 0.5% larger than those obtained from the X-ray spectra calculated from the aluminum half value layer and the tube voltage of incident X-rays. This study shows experimentally that the absorbed dose in a water-like phantom can be calculated with good accuracy by using the absorbed dose conversion factor obtained from the incident X-rays.

Open-access reader

About this research paper

What this paper is about

The diagnostic X-ray spectra in a water equivalent phantom have been measured. From these measured spectra, the absorbed dose conversion factors of water were derived. The primary X-ray spectra were also measured and the scattered X-ray spectra were calculated by subtraction. The measurements were made at the depths of 0, 5, 10, 15, and 20 cm in a 20 cm-thick phantom and at the X-ray tube voltages 60, 90, and 120 kV by using a small silicon diode detector. The radiation field size was 30×30 cm2 at the phantom surface. In the obtained spectra, the fraction of the scattered photon number is increased with the depth. The X-ray qualities of the spectra in the phantom were near the qualities of primary X-rays when the depth is 0 cm, and became near the qualities of scattered X-rays as the depth increases. The changes of the X-ray qualities due to the depth change were small; photon mean-energy changes were within 4.6 keV. The changes in the absorbed dose conversion factors were also small (within 0.68%). These conversion factors were 0.4–2.3% larger than those obtained from the effective energy of incident X-rays and only −0.3 to 0.5% larger than those obtained from the X-ray spectra calculated from the aluminum half value layer and the tube voltage of incident X-rays. This study shows experimentally that the absorbed dose in a water-like phantom can be calculated with good accuracy by using the absorbed dose conversion factor obtained from the incident X-rays.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The diagnostic X-ray spectra in a water equivalent phantom have been measured. From these measured spectra, the absorbed dose conversion factors of water were derived. The primary X-ray spectra were also measured and the scattered X-ray spectra were calculated by subtraction. The measurements were made at the depths of 0, 5, 10, 15, and 20 cm in a 20 cm-thick phantom and at the X-ray tube voltages 60, 90, and 120 kV by using a small silicon diode detector. The radiation field size was 30×30 cm2 at the phantom surface. In the obtained spectra, the fraction of the scattered photon number is increased with the depth. The X-ray qualities of the spectra in the phantom were near the qualities of primary X-rays when the depth is 0 cm, and became near the qualities of scattered X-rays as the depth increases. The changes of the X-ray qualities due to the depth change were small; photon mean-energy changes were within 4.6 keV. The changes in the absorbed dose conversion factors were also small (within 0.68%). These conversion factors were 0.4–2.3% larger than those obtained from the effective energy of incident X-rays and only −0.3 to 0.5% larger than those obtained from the X-ray spectra calculated from the aluminum half value layer and the tube voltage of incident X-rays. This study shows experimentally that the absorbed dose in a water-like phantom can be calculated with good accuracy by using the absorbed dose conversion factor obtained from the incident X-rays.

Key concepts: Imaging phantom, Spectral line, Absorbed dose, Irradiation, X-ray, Materials science, Photon, Radiation