Measured X-ray Spectra and Absorbed Dose Conversion Factors in a Water Equivalent Phantom Irradiated with Diagnostic X-rays
Kiyoshi Aoki
Abstract
Open-access reader
Kiyoshi Aoki
Abstract
Open-access reader
The diagnostic X-ray spectra in a water equivalent phantom have been measured. From these measured spectra, the absorbed dose conversion factors of water were derived. The primary X-ray spectra were also measured and the scattered X-ray spectra were calculated by subtraction. The measurements were made at the depths of 0, 5, 10, 15, and 20 cm in a 20 cm-thick phantom and at the X-ray tube voltages 60, 90, and 120 kV by using a small silicon diode detector. The radiation field size was 30×30 cm2 at the phantom surface. In the obtained spectra, the fraction of the scattered photon number is increased with the depth. The X-ray qualities of the spectra in the phantom were near the qualities of primary X-rays when the depth is 0 cm, and became near the qualities of scattered X-rays as the depth increases. The changes of the X-ray qualities due to the depth change were small; photon mean-energy changes were within 4.6 keV. The changes in the absorbed dose conversion factors were also small (within 0.68%). These conversion factors were 0.4–2.3% larger than those obtained from the effective energy of incident X-rays and only −0.3 to 0.5% larger than those obtained from the X-ray spectra calculated from the aluminum half value layer and the tube voltage of incident X-rays. This study shows experimentally that the absorbed dose in a water-like phantom can be calculated with good accuracy by using the absorbed dose conversion factor obtained from the incident X-rays.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The diagnostic X-ray spectra in a water equivalent phantom have been measured. From these measured spectra, the absorbed dose conversion factors of water were derived. The primary X-ray spectra were also measured and the scattered X-ray spectra were calculated by subtraction. The measurements were made at the depths of 0, 5, 10, 15, and 20 cm in a 20 cm-thick phantom and at the X-ray tube voltages 60, 90, and 120 kV by using a small silicon diode detector. The radiation field size was 30×30 cm2 at the phantom surface. In the obtained spectra, the fraction of the scattered photon number is increased with the depth. The X-ray qualities of the spectra in the phantom were near the qualities of primary X-rays when the depth is 0 cm, and became near the qualities of scattered X-rays as the depth increases. The changes of the X-ray qualities due to the depth change were small; photon mean-energy changes were within 4.6 keV. The changes in the absorbed dose conversion factors were also small (within 0.68%). These conversion factors were 0.4–2.3% larger than those obtained from the effective energy of incident X-rays and only −0.3 to 0.5% larger than those obtained from the X-ray spectra calculated from the aluminum half value layer and the tube voltage of incident X-rays. This study shows experimentally that the absorbed dose in a water-like phantom can be calculated with good accuracy by using the absorbed dose conversion factor obtained from the incident X-rays.
Key concepts: Imaging phantom, Spectral line, Absorbed dose, Irradiation, X-ray, Materials science, Photon, Radiation