2018•Radiation OncologyOpen access

Development of a PMMA phantom as a practical alternative for quality control of gamma knife® dosimetry

Jae Pil Chung, Young Min Seong, Tae‐Yeon Kim, Yona Choi, Tae Hoon Kim, Hyun Joon Choi, Chul Hee Min, Hamza Benmakhlouf, Kook Jin Chun, Hyun‐Tai Chung

Open full text 21 citations

Abstract

To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom. A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm 2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites. The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm 2 . The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60 Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm. A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.

Open-access reader

About this research paper

What this paper is about

To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom. A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm 2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites. The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm 2 . The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60 Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm. A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.

Why it matters

OpenAlex reports 21 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom. A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm 2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites. The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm 2 . The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60 Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm. A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.