THE RESPONSE OF A MULTI-ELEMENT TISSUE EQUIVALENT PROPORTIONAL COUNTER IN ACCELERATOR BASED HIGH-ALTITUDE NEUTRON FIELDS

Gloria M. Orchard, A.J. Waker

Open publisher page 0 citations

Abstract

The response of a multi-element tissue equivalent proportional counter (METEPC) was investigated at the high energy neutron facility EU-CERF, in Prevessin, France. This facility was established specifically to provide reference neutron fields typical of those found at high altitude. The METEPC measurements were conducted along with a commercially available spherical tissue equivalent proportional counter (TEPC) used for comparison. The measured microdosimetric spectra were analyzed and compared for various measurement locations, with substantially different neutron energy distributions resulting from the use of iron or concrete shielding. The absorbed dose rate measured by the two counters in the dominant lower-energy neutron field locations from the iron-shielded target was <27%. A significant difference in the range of 32-58% was observed between the two counters in the higher-energy neutron field produced by the concrete-shielded target. Interestingly, due to a combination of factors, including the geometry and structure of the METEPC, values of dose-equivalent derived from the METEPC and the TEPC were in closer agreement (within 30%) and are well within acceptable limits for neutron monitoring.

About this research paper

What this paper is about

The response of a multi-element tissue equivalent proportional counter (METEPC) was investigated at the high energy neutron facility EU-CERF, in Prevessin, France. This facility was established specifically to provide reference neutron fields typical of those found at high altitude. The METEPC measurements were conducted along with a commercially available spherical tissue equivalent proportional counter (TEPC) used for comparison. The measured microdosimetric spectra were analyzed and compared for various measurement locations, with substantially different neutron energy distributions resulting from the use of iron or concrete shielding. The absorbed dose rate measured by the two counters in the dominant lower-energy neutron field locations from the iron-shielded target was <27%. A significant difference in the range of 32-58% was observed between the two counters in the higher-energy neutron field produced by the concrete-shielded target. Interestingly, due to a combination of factors, including the geometry and structure of the METEPC, values of dose-equivalent derived from the METEPC and the TEPC were in closer agreement (within 30%) and are well within acceptable limits for neutron monitoring.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The response of a multi-element tissue equivalent proportional counter (METEPC) was investigated at the high energy neutron facility EU-CERF, in Prevessin, France. This facility was established specifically to provide reference neutron fields typical of those found at high altitude. The METEPC measurements were conducted along with a commercially available spherical tissue equivalent proportional counter (TEPC) used for comparison. The measured microdosimetric spectra were analyzed and compared for various measurement locations, with substantially different neutron energy distributions resulting from the use of iron or concrete shielding. The absorbed dose rate measured by the two counters in the dominant lower-energy neutron field locations from the iron-shielded target was <27%. A significant difference in the range of 32-58% was observed between the two counters in the higher-energy neutron field produced by the concrete-shielded target. Interestingly, due to a combination of factors, including the geometry and structure of the METEPC, values of dose-equivalent derived from the METEPC and the TEPC were in closer agreement (within 30%) and are well within acceptable limits for neutron monitoring.

Key concepts: Proportional counter, Neutron, Shielded cable, Equivalent dose, Electromagnetic shielding, Physics, Neutron temperature, Nuclear physics

Related papers

Back to paper searchBrowse research topicsOriginal source
THE RESPONSE OF A MULTI-ELEMENT TISSUE EQUIVALENT PROPORTIONAL COUNTER IN ACCELERATOR BASED HIGH-ALTITUDE NEUTRON FIELDS — Research Paper | ScholarLens