AN IN VITRO EVALUATION OF THE ACCURACY OF ROOT ZX ELECTRONIC APEX LOCATOR

Daehoon Kang, Kwang-Hee Chung, Soo‐Han Yoon, Kwang‐Shik Bae

Open publisher page 1 citations

Abstract

The purpose of this study was to evaluate the in vitro accuracy of Root ZX(Morita Co., Japan) which is the ratio type electronic apex locator. The 86 extracted human palatal roots of maxillary molar with fully formed apices were used. File lengths with the file tip just visible at the foramen were compared to those measured with Root ZX. For length measuring with Root ZX, saline test model with which the apical 1/3 of each root was submerged into normal saline were designed. The root canal lengths were determined with Root ZX and the radiographs were taken with a file in the canal. The distances from file tips of Root ZX lengths to apecies in radiographs also were measured with Profile projector PJ311(Mitutoyo Co., Japan). The results were as follows : 1. The root canal length determined with electronic apex locator was shorter than the length with visual measurement. 2. The file tip of Root ZX lengths was located at away from the apex in radiograph. 3. The accuracy of the Root ZX was 79.1% within 0.5mm of visual working length and 96.5% within 1.0mm.

About this research paper

What this paper is about

The purpose of this study was to evaluate the in vitro accuracy of Root ZX(Morita Co., Japan) which is the ratio type electronic apex locator. The 86 extracted human palatal roots of maxillary molar with fully formed apices were used. File lengths with the file tip just visible at the foramen were compared to those measured with Root ZX. For length measuring with Root ZX, saline test model with which the apical 1/3 of each root was submerged into normal saline were designed. The root canal lengths were determined with Root ZX and the radiographs were taken with a file in the canal. The distances from file tips of Root ZX lengths to apecies in radiographs also were measured with Profile projector PJ311(Mitutoyo Co., Japan). The results were as follows : 1. The root canal length determined with electronic apex locator was shorter than the length with visual measurement. 2. The file tip of Root ZX lengths was located at away from the apex in radiograph. 3. The accuracy of the Root ZX was 79.1% within 0.5mm of visual working length and 96.5% within 1.0mm.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The purpose of this study was to evaluate the in vitro accuracy of Root ZX(Morita Co., Japan) which is the ratio type electronic apex locator. The 86 extracted human palatal roots of maxillary molar with fully formed apices were used. File lengths with the file tip just visible at the foramen were compared to those measured with Root ZX. For length measuring with Root ZX, saline test model with which the apical 1/3 of each root was submerged into normal saline were designed. The root canal lengths were determined with Root ZX and the radiographs were taken with a file in the canal. The distances from file tips of Root ZX lengths to apecies in radiographs also were measured with Profile projector PJ311(Mitutoyo Co., Japan). The results were as follows : 1. The root canal length determined with electronic apex locator was shorter than the length with visual measurement. 2. The file tip of Root ZX lengths was located at away from the apex in radiograph. 3. The accuracy of the Root ZX was 79.1% within 0.5mm of visual working length and 96.5% within 1.0mm.

Key concepts: Apex (geometry), Root canal, Apical foramen, Root (linguistics), Mathematics, Materials science, Horticulture, Dentistry

Related papers

Back to paper searchBrowse research topicsOriginal source
AN IN VITRO EVALUATION OF THE ACCURACY OF ROOT ZX ELECTRONIC APEX LOCATOR — Research Paper | ScholarLens