1999International Endodontic JournalRequires access

In vitro evaluation of the reliability of the Root ZX electronic apex locator

Hani Ounsi, Alfred Naaman

Open publisher page 95 citations

Abstract

AIM: An experimental comparison of the Root ZX apex locator with the real measurement of the root canal length was carried out using a saline gel to simulate the periodontium. METHODOLOGY: Lengths were taken when the needle reached the 0.5 mark and the Apex mark on the meter. These measurements were then compared with a calculated reference length representing the real length of the root canal. Reliability was assessed using Student's t-test. Precision was assessed using the mean and standard deviation of differences. RESULTS: The Root ZX gave measurements within a range of 0.5 mm in 84.72% of the cases. The intra-operator and inter-operator variabilities were not statistically significant (P > 0.05). CONCLUSIONS: The Root ZX is not capable of detecting the '0.5 mm from the foramen' position and thus, should only be used to detect the foramen (major diameter).

About this research paper

What this paper is about

AIM: An experimental comparison of the Root ZX apex locator with the real measurement of the root canal length was carried out using a saline gel to simulate the periodontium. METHODOLOGY: Lengths were taken when the needle reached the 0.5 mark and the Apex mark on the meter. These measurements were then compared with a calculated reference length representing the real length of the root canal. Reliability was assessed using Student's t-test. Precision was assessed using the mean and standard deviation of differences. RESULTS: The Root ZX gave measurements within a range of 0.5 mm in 84.72% of the cases. The intra-operator and inter-operator variabilities were not statistically significant (P > 0.05). CONCLUSIONS: The Root ZX is not capable of detecting the '0.5 mm from the foramen' position and thus, should only be used to detect the foramen (major diameter).

Why it matters

OpenAlex reports 95 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

AIM: An experimental comparison of the Root ZX apex locator with the real measurement of the root canal length was carried out using a saline gel to simulate the periodontium. METHODOLOGY: Lengths were taken when the needle reached the 0.5 mark and the Apex mark on the meter. These measurements were then compared with a calculated reference length representing the real length of the root canal. Reliability was assessed using Student's t-test. Precision was assessed using the mean and standard deviation of differences. RESULTS: The Root ZX gave measurements within a range of 0.5 mm in 84.72% of the cases. The intra-operator and inter-operator variabilities were not statistically significant (P > 0.05). CONCLUSIONS: The Root ZX is not capable of detecting the '0.5 mm from the foramen' position and thus, should only be used to detect the foramen (major diameter).

Key concepts: Apical foramen, Apex (geometry), Root canal, Reliability (semiconductor), Root (linguistics), Mathematics, Medicine, Orthodontics

Related papers

Back to paper searchBrowse research topicsOriginal source
In vitro evaluation of the reliability of the Root ZX electronic apex locator — Research Paper | ScholarLens