1976•Journal of Nuclear Medicine TechnologyRequires access

Quality Control Procedure for Pipetting Systems

JANET M. MARKS, Aline Zimmer, Edward A. Silverstein, Richard A. Holmes

Open publisher page 0 citations

Abstract

Our laboratory has developed a simple and rapid radionuclide procedure to make evaluations of the reproducibility of pipettes. Five semiautomatic and one automatic pipette were evaluated by delivering diluted 125I-human serum albumin aqueous or pooled serum samples. Each sample was counted in an autogamma counter. The results were statistically analyzed by calculating means, standard deviations, and coefficients of variation. The automatic pipette and one semiautomatic pipette demonstrated low coefficients of variation. All other pipettes evaluated showed inconsistency in the delivered volumes. The radionuclide method, as outlined in this study, can easily be incorporated as a quality control procedure for pipetting systems in any nuclear medicine laboratory.

About this research paper

What this paper is about

Our laboratory has developed a simple and rapid radionuclide procedure to make evaluations of the reproducibility of pipettes. Five semiautomatic and one automatic pipette were evaluated by delivering diluted 125I-human serum albumin aqueous or pooled serum samples. Each sample was counted in an autogamma counter. The results were statistically analyzed by calculating means, standard deviations, and coefficients of variation. The automatic pipette and one semiautomatic pipette demonstrated low coefficients of variation. All other pipettes evaluated showed inconsistency in the delivered volumes. The radionuclide method, as outlined in this study, can easily be incorporated as a quality control procedure for pipetting systems in any nuclear medicine laboratory.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Our laboratory has developed a simple and rapid radionuclide procedure to make evaluations of the reproducibility of pipettes. Five semiautomatic and one automatic pipette were evaluated by delivering diluted 125I-human serum albumin aqueous or pooled serum samples. Each sample was counted in an autogamma counter. The results were statistically analyzed by calculating means, standard deviations, and coefficients of variation. The automatic pipette and one semiautomatic pipette demonstrated low coefficients of variation. All other pipettes evaluated showed inconsistency in the delivered volumes. The radionuclide method, as outlined in this study, can easily be incorporated as a quality control procedure for pipetting systems in any nuclear medicine laboratory.

Key concepts: Pipette, Reproducibility, Biomedical engineering, Coefficient of variation, Computer science, Chromatography, Nuclear medicine, Chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
Quality Control Procedure for Pipetting Systems — Research Paper | ScholarLens