[The eccentriplate device on the Dideco Vivacell cell separator for the collection of platelets].
G. Andreu, Alexandrine Gautheron, A Nicod-Tran, C Lecrubier, M Samama
Abstract
G. Andreu, Alexandrine Gautheron, A Nicod-Tran, C Lecrubier, M Samama
Abstract
When adapted to the Dideco Vivacell cell separator, the eccentriplate system allows preparation of leukocyte-poor platelet concentrates. Results of its use on 84 procedures showed that a mean of 4.51 x 10(11) platelets were removed in less than 2 hours with a particularly low level of leukocyte contamination: 2.6 x 10(8). Study of platelet function showed them to be altered more by eccentriplate than by conventional removal with the same apparatus. Analysis of recirculation in vivo corroborated the functional anomalies observed with all aggregating agents (with the exception of arachidonate). These anomalies were associated with a quantity of platelets collected at the lower limit of French norms: 30% of the concentrates contained less than 4 x 10(11) platelets. These findings indicate the need for improvement in the eccentriplate system.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
When adapted to the Dideco Vivacell cell separator, the eccentriplate system allows preparation of leukocyte-poor platelet concentrates. Results of its use on 84 procedures showed that a mean of 4.51 x 10(11) platelets were removed in less than 2 hours with a particularly low level of leukocyte contamination: 2.6 x 10(8). Study of platelet function showed them to be altered more by eccentriplate than by conventional removal with the same apparatus. Analysis of recirculation in vivo corroborated the functional anomalies observed with all aggregating agents (with the exception of arachidonate). These anomalies were associated with a quantity of platelets collected at the lower limit of French norms: 30% of the concentrates contained less than 4 x 10(11) platelets. These findings indicate the need for improvement in the eccentriplate system.
Key concepts: Platelet, Medicine, Separator (oil production), In vivo, Immunology, Genetics, Biology, Physics