1994PubMedRequires access

Modification of an intrauterine shunt catheter to reduce the incidence of spontaneous displacement.

Douglas Richards, Gregory J. Locksmith, M K Yancey

Open publisher page 0 citations

Abstract

OBJECTIVE: To determine which of four types of catheters would be the least likely to become displaced spontaneously when employed as a fetoamniotic shunt, by testing the force required to remove the catheters from a lamb's bladder. METHODS: The catheter designs tested were unfeathered straight, unfeathered pig-tailed, superficially feathered, and deeply feathered. The catheters were placed percutaneously into the urinary bladders of recently euthanized newborn lambs. The force required to withdraw the catheters was determined, employing nine measurements for each catheter design. RESULTS: The mean force required to remove the straight catheter was 12 g (95% confidence interval [CI] 2-21 g), the pig-tailed catheter 22 g (95% CI 13-32 g), and the superficially feathered catheter 149 g (95% CI 139-160 g). Extracting the deeply feathered catheter consistently required a force greater than 300 g, which was the upper limit of detection for our instruments. CONCLUSION: Feathering the catheter increases the force required to extract it from a fetal viscus. We believe that use of feathered catheters will significantly reduce the incidence of unwanted spontaneous displacement.

About this research paper

What this paper is about

OBJECTIVE: To determine which of four types of catheters would be the least likely to become displaced spontaneously when employed as a fetoamniotic shunt, by testing the force required to remove the catheters from a lamb's bladder. METHODS: The catheter designs tested were unfeathered straight, unfeathered pig-tailed, superficially feathered, and deeply feathered. The catheters were placed percutaneously into the urinary bladders of recently euthanized newborn lambs. The force required to withdraw the catheters was determined, employing nine measurements for each catheter design. RESULTS: The mean force required to remove the straight catheter was 12 g (95% confidence interval [CI] 2-21 g), the pig-tailed catheter 22 g (95% CI 13-32 g), and the superficially feathered catheter 149 g (95% CI 139-160 g). Extracting the deeply feathered catheter consistently required a force greater than 300 g, which was the upper limit of detection for our instruments. CONCLUSION: Feathering the catheter increases the force required to extract it from a fetal viscus. We believe that use of feathered catheters will significantly reduce the incidence of unwanted spontaneous displacement.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: To determine which of four types of catheters would be the least likely to become displaced spontaneously when employed as a fetoamniotic shunt, by testing the force required to remove the catheters from a lamb's bladder. METHODS: The catheter designs tested were unfeathered straight, unfeathered pig-tailed, superficially feathered, and deeply feathered. The catheters were placed percutaneously into the urinary bladders of recently euthanized newborn lambs. The force required to withdraw the catheters was determined, employing nine measurements for each catheter design. RESULTS: The mean force required to remove the straight catheter was 12 g (95% confidence interval [CI] 2-21 g), the pig-tailed catheter 22 g (95% CI 13-32 g), and the superficially feathered catheter 149 g (95% CI 139-160 g). Extracting the deeply feathered catheter consistently required a force greater than 300 g, which was the upper limit of detection for our instruments. CONCLUSION: Feathering the catheter increases the force required to extract it from a fetal viscus. We believe that use of feathered catheters will significantly reduce the incidence of unwanted spontaneous displacement.

Key concepts: Catheter, Medicine, Shunt (medical), Surgery, Urinary catheter, Confidence interval, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Modification of an intrauterine shunt catheter to reduce the incidence of spontaneous displacement. — Research Paper | ScholarLens