[Determination of tryptophan in human serum by high-performance liquid chromatography].
AI-QIONG TANG, Q Meng, J Wang
Abstract
AI-QIONG TANG, Q Meng, J Wang
Abstract
In this paper a high-performance liquid chromatographic(HPLC) method has been developed to determine the concentration of tryptophan(Trp) in human serum. Serum sample was precipitated with trichloroacetic acid solution(0.6 mol.L-1) to remove protein and centrifuged and assayed by HPLC. The operating conditions were NOVA-PAK C18 column(3.9 mm x 150 mm), KH2PO4(0.1 mol.L-1) mobile phase with flow rate of 0.7 ml.min-1, and detective wave-length at 295 nm. The linear range was 1.223-2447.381 mumol.L-1. The detection limit was 0.612 mumol.L-1. The average recovery was 96.3%. Intra-assay CV is less than 4% and inter-assay less than 5%. The concentration of Trp in normal adult serum was 46.603 +/- 5.108 mumol.L-1. The detection time of serum Trp was within 5 min. The results suggest that this method is simple, fast, accurate, and convenient.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
In this paper a high-performance liquid chromatographic(HPLC) method has been developed to determine the concentration of tryptophan(Trp) in human serum. Serum sample was precipitated with trichloroacetic acid solution(0.6 mol.L-1) to remove protein and centrifuged and assayed by HPLC. The operating conditions were NOVA-PAK C18 column(3.9 mm x 150 mm), KH2PO4(0.1 mol.L-1) mobile phase with flow rate of 0.7 ml.min-1, and detective wave-length at 295 nm. The linear range was 1.223-2447.381 mumol.L-1. The detection limit was 0.612 mumol.L-1. The average recovery was 96.3%. Intra-assay CV is less than 4% and inter-assay less than 5%. The concentration of Trp in normal adult serum was 46.603 +/- 5.108 mumol.L-1. The detection time of serum Trp was within 5 min. The results suggest that this method is simple, fast, accurate, and convenient.
Key concepts: Trichloroacetic acid, Chromatography, High-performance liquid chromatography, Detection limit, Chemistry, Tryptophan, Serum concentration, Linear range