Application of automatic hematology analyzer Sysmex XT-4000i in cells counting for ascite fluid

Wenjing Wang, Xiuying Zhao, Guo Hong-yan

Open publisher page 0 citations

Abstract

Objective To evaluate the application of body fluid mode in automatic hematology analyzer Sysmex XT-4000i (RET channel) in ascite fluid cells counting. Methods One hundred fresh ascite fluid specimens were used for cell counting by automatic hematology analyzer Sysmex XT-4000i (automatic method) and improved Neubauer counting board (manual method) respectively. The results were compared by statistics. Results A good correlation was found between automatic method and manual method. There was no significant difference of the results in leukocyte[(1.403±5.527)×109/L vs. (0.858±2.740) ×109/L], mononuclear cells[(0.430±1.172) ×109/L vs. (0.343±0.832) ×109/L] and multinucleated cells[(0.969±5.011) ×109/L vs. (0.515±2.371) ×109/L] counting by automatic method and manual method (P 0.05). The coefficient variation(CV%) of the automatic method was higher than manual method.The carry over rates of total cells, erythrocytes, leukocytes, mononuclear cells were 0.037%, 0.000%, 0.161%, 0.070%, while the carry over rates of multinucleated cells was 5.633%. Conclusion Sysmex XT-4000i can be applied in cells counting for routinely clinical ascite fluid detection.

About this research paper

What this paper is about

Objective To evaluate the application of body fluid mode in automatic hematology analyzer Sysmex XT-4000i (RET channel) in ascite fluid cells counting. Methods One hundred fresh ascite fluid specimens were used for cell counting by automatic hematology analyzer Sysmex XT-4000i (automatic method) and improved Neubauer counting board (manual method) respectively. The results were compared by statistics. Results A good correlation was found between automatic method and manual method. There was no significant difference of the results in leukocyte[(1.403±5.527)×109/L vs. (0.858±2.740) ×109/L], mononuclear cells[(0.430±1.172) ×109/L vs. (0.343±0.832) ×109/L] and multinucleated cells[(0.969±5.011) ×109/L vs. (0.515±2.371) ×109/L] counting by automatic method and manual method (P 0.05). The coefficient variation(CV%) of the automatic method was higher than manual method.The carry over rates of total cells, erythrocytes, leukocytes, mononuclear cells were 0.037%, 0.000%, 0.161%, 0.070%, while the carry over rates of multinucleated cells was 5.633%. Conclusion Sysmex XT-4000i can be applied in cells counting for routinely clinical ascite fluid detection.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the application of body fluid mode in automatic hematology analyzer Sysmex XT-4000i (RET channel) in ascite fluid cells counting. Methods One hundred fresh ascite fluid specimens were used for cell counting by automatic hematology analyzer Sysmex XT-4000i (automatic method) and improved Neubauer counting board (manual method) respectively. The results were compared by statistics. Results A good correlation was found between automatic method and manual method. There was no significant difference of the results in leukocyte[(1.403±5.527)×109/L vs. (0.858±2.740) ×109/L], mononuclear cells[(0.430±1.172) ×109/L vs. (0.343±0.832) ×109/L] and multinucleated cells[(0.969±5.011) ×109/L vs. (0.515±2.371) ×109/L] counting by automatic method and manual method (P 0.05). The coefficient variation(CV%) of the automatic method was higher than manual method.The carry over rates of total cells, erythrocytes, leukocytes, mononuclear cells were 0.037%, 0.000%, 0.161%, 0.070%, while the carry over rates of multinucleated cells was 5.633%. Conclusion Sysmex XT-4000i can be applied in cells counting for routinely clinical ascite fluid detection.

Key concepts: Hematology analyzer, Cell counting, Medicine, Coefficient of variation, Nuclear medicine, Pathology, Internal medicine, Chromatography

Related papers

Back to paper searchBrowse research topicsOriginal source
Application of automatic hematology analyzer Sysmex XT-4000i in cells counting for ascite fluid — Research Paper | ScholarLens