A Preliminary Study of Dual-source CT Coronary Angiography in Low Kilovoltage Setting
Changsheng Zhou
Abstract
Changsheng Zhou
Abstract
Objective To evaluate the feasibility of low tube voltage techniques in CT coronary angiography(CTCA) with dual-source CT(DSCT)scanner.Materials and Methods 60 patients undergoing CTCA were enrolled into this study,who were divided into conventional voltage group(120 kVp,A group,n=30)and low tube voltage group(100 kVp,B group,n=30).Image quality,contrast noise ratio(CNR),volume CT dose index(CTDIvol),and effective dose(ED) of two groups were evaluated respectively.Results The mean image quality score of RCA,LAD,and CX was 3.77±0.50,3.73±0.52,3.80±0.48 for group A and 3.50±0.63,3.53±0.60,3.50±0.57 for group B,respectively;the mean image CNR was 20.33±5.14 21.56±5.61 for group A and group B,respectively.There was no significant difference in the image quality score and the mean image noise between the two groups(t=-0.88,P0.05),the average CTDIvol in A group and B group was(62.24±10.12)mGy and(36.50±8.45)mGy,respectively(t=10.7,P0.05);and the mean ED was(17.63±3.75)mSv and(10.03±2.23)mSv,respectively(t=9.55,P0.05).Conclusion Low tube voltage CTCA can reduce radiation dose to the patient,and meet the demand of the clinic diagnosis.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the feasibility of low tube voltage techniques in CT coronary angiography(CTCA) with dual-source CT(DSCT)scanner.Materials and Methods 60 patients undergoing CTCA were enrolled into this study,who were divided into conventional voltage group(120 kVp,A group,n=30)and low tube voltage group(100 kVp,B group,n=30).Image quality,contrast noise ratio(CNR),volume CT dose index(CTDIvol),and effective dose(ED) of two groups were evaluated respectively.Results The mean image quality score of RCA,LAD,and CX was 3.77±0.50,3.73±0.52,3.80±0.48 for group A and 3.50±0.63,3.53±0.60,3.50±0.57 for group B,respectively;the mean image CNR was 20.33±5.14 21.56±5.61 for group A and group B,respectively.There was no significant difference in the image quality score and the mean image noise between the two groups(t=-0.88,P0.05),the average CTDIvol in A group and B group was(62.24±10.12)mGy and(36.50±8.45)mGy,respectively(t=10.7,P0.05);and the mean ED was(17.63±3.75)mSv and(10.03±2.23)mSv,respectively(t=9.55,P0.05).Conclusion Low tube voltage CTCA can reduce radiation dose to the patient,and meet the demand of the clinic diagnosis.
Key concepts: Medicine, Image quality, Nuclear medicine, Radiation dose, Image noise, Contrast-to-noise ratio, Effective dose (radiation), Coronary angiography