Clinical application of calibration for automated hematology analyzers with fresh blood

Zhang Hon

Open publisher page 0 citations

Abstract

Objective To calibrate hematology analyzers with fresh blood to make their parameters comparable. Methods Fresh blood assigned by hematology analyzer Sysmex XT-1800i as normative measurement system was used to calibrate two Sysmex KX-21 analyzers and one KX-21N analyzer respectively.Results In the parameters calibrated for hematology analyzers,there was about 53.3%(8/15) over allowable bias before calibration but 13.3%(2/15) after calibration with fresh blood.Conclusions Parameters of hematology analyzers calibrated with fresh blood are comparable and coincident.

About this research paper

What this paper is about

Objective To calibrate hematology analyzers with fresh blood to make their parameters comparable. Methods Fresh blood assigned by hematology analyzer Sysmex XT-1800i as normative measurement system was used to calibrate two Sysmex KX-21 analyzers and one KX-21N analyzer respectively.Results In the parameters calibrated for hematology analyzers,there was about 53.3%(8/15) over allowable bias before calibration but 13.3%(2/15) after calibration with fresh blood.Conclusions Parameters of hematology analyzers calibrated with fresh blood are comparable and coincident.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To calibrate hematology analyzers with fresh blood to make their parameters comparable. Methods Fresh blood assigned by hematology analyzer Sysmex XT-1800i as normative measurement system was used to calibrate two Sysmex KX-21 analyzers and one KX-21N analyzer respectively.Results In the parameters calibrated for hematology analyzers,there was about 53.3%(8/15) over allowable bias before calibration but 13.3%(2/15) after calibration with fresh blood.Conclusions Parameters of hematology analyzers calibrated with fresh blood are comparable and coincident.

Key concepts: Hematology analyzer, Hematology, Calibration, Medicine, Internal medicine, Medical physics, Nuclear medicine, Mathematics

Related papers

Back to paper searchBrowse research topicsOriginal source
Clinical application of calibration for automated hematology analyzers with fresh blood — Research Paper | ScholarLens