Impact of iterative reconstruction technique and radiation dose on low-dose chest CT image quality

Ya Xu

Open publisher page 0 citations

Abstract

Objective To evaluate the impact of iterative reconstruction technique(iDOSE4)and radiation dose on image noise and contrast to noise ratio(CNR)of low-dose chest CT using a chest phantom.Methods A 64-detector spiral CT was used to scan a chest phantom with different tube current(30mAs,50mAs,70mAs)and tube voltage(80kV,100kV).CT original images were reconstructed with filtered back projection(FBP),iDOSE4 L3and L6algorithm,respectively.Then two radiologists measured CT value of ROI on reconstruction images,and average noise and CNR were calculated.Repeated measure analysis of variance was used to compare CT values and image quality among different radiation doses and reconstruction algorithm.Results There was no statistical difference of CT value of air and thorax lung field homogeneous among different radiation doses and reconstruction algorithms(all P0.05).With the increase of radiation dose,the average noise decreased and CNR increased gradually in the same reconstruction algorithm.In the same radiation dose,the average noise of images with FBP,iDOSE4 L3and L6reconstruction algorithm decreased and CNR increased gradually.Conclusion Changes of radiation dose and(or)reconstruction algorithm can not affect measurement accuracy of CT value.iDOSE4 algorithm can reduce average noise and improve image quality of low-dose CT image.

About this research paper

What this paper is about

Objective To evaluate the impact of iterative reconstruction technique(iDOSE4)and radiation dose on image noise and contrast to noise ratio(CNR)of low-dose chest CT using a chest phantom.Methods A 64-detector spiral CT was used to scan a chest phantom with different tube current(30mAs,50mAs,70mAs)and tube voltage(80kV,100kV).CT original images were reconstructed with filtered back projection(FBP),iDOSE4 L3and L6algorithm,respectively.Then two radiologists measured CT value of ROI on reconstruction images,and average noise and CNR were calculated.Repeated measure analysis of variance was used to compare CT values and image quality among different radiation doses and reconstruction algorithm.Results There was no statistical difference of CT value of air and thorax lung field homogeneous among different radiation doses and reconstruction algorithms(all P0.05).With the increase of radiation dose,the average noise decreased and CNR increased gradually in the same reconstruction algorithm.In the same radiation dose,the average noise of images with FBP,iDOSE4 L3and L6reconstruction algorithm decreased and CNR increased gradually.Conclusion Changes of radiation dose and(or)reconstruction algorithm can not affect measurement accuracy of CT value.iDOSE4 algorithm can reduce average noise and improve image quality of low-dose CT image.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the impact of iterative reconstruction technique(iDOSE4)and radiation dose on image noise and contrast to noise ratio(CNR)of low-dose chest CT using a chest phantom.Methods A 64-detector spiral CT was used to scan a chest phantom with different tube current(30mAs,50mAs,70mAs)and tube voltage(80kV,100kV).CT original images were reconstructed with filtered back projection(FBP),iDOSE4 L3and L6algorithm,respectively.Then two radiologists measured CT value of ROI on reconstruction images,and average noise and CNR were calculated.Repeated measure analysis of variance was used to compare CT values and image quality among different radiation doses and reconstruction algorithm.Results There was no statistical difference of CT value of air and thorax lung field homogeneous among different radiation doses and reconstruction algorithms(all P0.05).With the increase of radiation dose,the average noise decreased and CNR increased gradually in the same reconstruction algorithm.In the same radiation dose,the average noise of images with FBP,iDOSE4 L3and L6reconstruction algorithm decreased and CNR increased gradually.Conclusion Changes of radiation dose and(or)reconstruction algorithm can not affect measurement accuracy of CT value.iDOSE4 algorithm can reduce average noise and improve image quality of low-dose CT image.

Key concepts: Medicine, Imaging phantom, Iterative reconstruction, Image noise, Image quality, Nuclear medicine, Reconstruction algorithm, Radiation dose

Related papers

Back to paper searchBrowse research topicsOriginal source
Impact of iterative reconstruction technique and radiation dose on low-dose chest CT image quality — Research Paper | ScholarLens