The application of Sysmex XE-2100 automatic hematology analyzer in the determination of leukocyte differential count
LI Zao-rong
Abstract
LI Zao-rong
Abstract
Objective:To evaluate the accuracy of Sysmex XE-2100 automatic hematology analyzer in the determination of leukocyte differential count,and to provide basis for choosing test methods in clinic.Methods:A total of 60 samples of venous blood were determined for leukocyte differential count with Sysmex XE-2100 automatic hematology analyzer and microscope respectively,with 10 times for each sample,and the precision was analyzed,then the results of leukocyte differential count of these two methods were underwent correlation analysis.Results:Both the within-batch precision and between-run precision of Sysmex XE-2100 automatic hematology analyzer were in the allowed ranges.The results of leukocyte differential count with both Sysmex XE-2100 automatic hematology analyzer and microscope showed a good correlation,and the coefficient correlations of neutrophile granulocytes,leukomonocytes,monocytes,eosinophilic granulocytes and basophilic granulocytes were 0.9238,0.9764,0.8230,0.8862 and 0.6047 respectively(all P0.05).Conclusion:Sysmex XE-2100 automatic hematology analyzer can rapidly and accurately determine leukocyte differential count,which has a good repeatability and is suitable for the determination of a large number of samples,thus it is an ideal analyzer for the tests in clinic.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To evaluate the accuracy of Sysmex XE-2100 automatic hematology analyzer in the determination of leukocyte differential count,and to provide basis for choosing test methods in clinic.Methods:A total of 60 samples of venous blood were determined for leukocyte differential count with Sysmex XE-2100 automatic hematology analyzer and microscope respectively,with 10 times for each sample,and the precision was analyzed,then the results of leukocyte differential count of these two methods were underwent correlation analysis.Results:Both the within-batch precision and between-run precision of Sysmex XE-2100 automatic hematology analyzer were in the allowed ranges.The results of leukocyte differential count with both Sysmex XE-2100 automatic hematology analyzer and microscope showed a good correlation,and the coefficient correlations of neutrophile granulocytes,leukomonocytes,monocytes,eosinophilic granulocytes and basophilic granulocytes were 0.9238,0.9764,0.8230,0.8862 and 0.6047 respectively(all P0.05).Conclusion:Sysmex XE-2100 automatic hematology analyzer can rapidly and accurately determine leukocyte differential count,which has a good repeatability and is suitable for the determination of a large number of samples,thus it is an ideal analyzer for the tests in clinic.
Key concepts: Hematology analyzer, Repeatability, Spectrum analyzer, Medicine, Leukocyte Counts, Hematology, Coefficient of variation, Pathology