2004Journal of Tropical MedicineRequires access

Evaluation on Blood Platelet Count by Sysmex XE-2100 Hematology Analyzer

Xiao Wang-xian

Open publisher page 0 citations

Abstract

Objective To evaluate the accuracy of Blood Platelet Count by Sysm ex XE-2100 Hematology Analyzer. Methods The precision, linear range and contami nate rate of blood platelet count were tested by Sysmex XE-2100 Hematology Anal yzer; 150 specimens were counted by Sysmex XE-2100 Hematology Analyzer and comp ared with those under microscopy, then to analyze the correlation. Results The p recision, linear range and contaminate rate of blood platelet count are in the a llowable range. There was significant correlation in the results between two met hods, the correlation coefficient of PLT-O and PLT-I were 0.9937 and 0.9854 (P 0.01). Conclusions Blood platelet count by Sysmex XE-2100 Hematology Analyzer is quick, convenient and accurate, it is satisfied with needs of blood platelet count in the clinical laboratory.

About this research paper

What this paper is about

Objective To evaluate the accuracy of Blood Platelet Count by Sysm ex XE-2100 Hematology Analyzer. Methods The precision, linear range and contami nate rate of blood platelet count were tested by Sysmex XE-2100 Hematology Anal yzer; 150 specimens were counted by Sysmex XE-2100 Hematology Analyzer and comp ared with those under microscopy, then to analyze the correlation. Results The p recision, linear range and contaminate rate of blood platelet count are in the a llowable range. There was significant correlation in the results between two met hods, the correlation coefficient of PLT-O and PLT-I were 0.9937 and 0.9854 (P 0.01). Conclusions Blood platelet count by Sysmex XE-2100 Hematology Analyzer is quick, convenient and accurate, it is satisfied with needs of blood platelet count in the clinical laboratory.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the accuracy of Blood Platelet Count by Sysm ex XE-2100 Hematology Analyzer. Methods The precision, linear range and contami nate rate of blood platelet count were tested by Sysmex XE-2100 Hematology Anal yzer; 150 specimens were counted by Sysmex XE-2100 Hematology Analyzer and comp ared with those under microscopy, then to analyze the correlation. Results The p recision, linear range and contaminate rate of blood platelet count are in the a llowable range. There was significant correlation in the results between two met hods, the correlation coefficient of PLT-O and PLT-I were 0.9937 and 0.9854 (P 0.01). Conclusions Blood platelet count by Sysmex XE-2100 Hematology Analyzer is quick, convenient and accurate, it is satisfied with needs of blood platelet count in the clinical laboratory.

Key concepts: Hematology analyzer, Hematology, Platelet, Medicine, Internal medicine, Spectrum analyzer, Linear correlation, Reference range