Performance Evaluation of Sysmex XS-800i Automated Hematology Analyzer
Aijun Huang
Abstract
Aijun Huang
Abstract
Objective To evaluate the function of the Sysmex XS-800i automated hematology analyzer.Methods Precision,linearity,carrier rate and repetition were analyzed.Blood samples from 60 healthy people were tested by Sysmex XS-800i and the Sysmex 2100 respectively.Four indexes including WBC,RBC,Hb and PLT from two tests were compared.The results differentiated by Sysmex XS-800i were compared with those by handwork.Results Precision,linearity,carrier rate and repetition studies of Sysmex XS-800i were all in the allowable range(supplied by factory authorities).The Statistics results of WBC,RBC,Hb and PLT tested by Sysmex XS-800i showed no significant change in comparison to those by Sysmex 2100(P0.05).The results differentiated by Sysmex XS-800i showed excellent correlation with those by handwork.Conclusion The performance of the Sysmex XS-800i automated hematology analyzer is satisfied.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the function of the Sysmex XS-800i automated hematology analyzer.Methods Precision,linearity,carrier rate and repetition were analyzed.Blood samples from 60 healthy people were tested by Sysmex XS-800i and the Sysmex 2100 respectively.Four indexes including WBC,RBC,Hb and PLT from two tests were compared.The results differentiated by Sysmex XS-800i were compared with those by handwork.Results Precision,linearity,carrier rate and repetition studies of Sysmex XS-800i were all in the allowable range(supplied by factory authorities).The Statistics results of WBC,RBC,Hb and PLT tested by Sysmex XS-800i showed no significant change in comparison to those by Sysmex 2100(P0.05).The results differentiated by Sysmex XS-800i showed excellent correlation with those by handwork.Conclusion The performance of the Sysmex XS-800i automated hematology analyzer is satisfied.
Key concepts: Hematology analyzer, Medicine, Internal medicine