Determination of cadmium by ethanol-ammonium sulfate two-phase extraction combined with flame atomic absorption spectrometry
Hou Yan-min
Abstract
Hou Yan-min
Abstract
The ethanol-ammonium sulfate two-phase extraction system for cadmium was developed.The main influencing factors for the extraction of cadmium by ethanol-ammonium sulfate two-phase system were investigated with flame atomic absorption spectrometry(FAAS).Moreover,the extraction mechanism was preliminarily discussed by Fourier transform infrared spectrometry(FT-IR).The results showed that Cd2+ could form stable purple complex with I-and crystal violet in ethanol-ammonium sulfate two-phase system at pH 5.0-6.0.Under the optimal extraction conditions,cadmium could be well separated from other common metal ions.The extraction rate was up to 100 %.The linear range of calibration curve was 0-0.50 μg/mL,with the linear correlation coefficient of 0.999 3 and the detection limit of 0.012 μg/mL.The proposed method was applied to the determination of cadmium in environmental water samples.The relative standard deviations(RSD) were 2.8 %-3.6 %,and the recoveries were higher than 95%.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The ethanol-ammonium sulfate two-phase extraction system for cadmium was developed.The main influencing factors for the extraction of cadmium by ethanol-ammonium sulfate two-phase system were investigated with flame atomic absorption spectrometry(FAAS).Moreover,the extraction mechanism was preliminarily discussed by Fourier transform infrared spectrometry(FT-IR).The results showed that Cd2+ could form stable purple complex with I-and crystal violet in ethanol-ammonium sulfate two-phase system at pH 5.0-6.0.Under the optimal extraction conditions,cadmium could be well separated from other common metal ions.The extraction rate was up to 100 %.The linear range of calibration curve was 0-0.50 μg/mL,with the linear correlation coefficient of 0.999 3 and the detection limit of 0.012 μg/mL.The proposed method was applied to the determination of cadmium in environmental water samples.The relative standard deviations(RSD) were 2.8 %-3.6 %,and the recoveries were higher than 95%.
Key concepts: Chemistry, Extraction (chemistry), Cadmium, Detection limit, Ammonium sulfate, Calibration curve, Analytical Chemistry (journal), Atomic absorption spectroscopy