Determination of trace mercury in geochemical samples by cold vapor atomic fluorescence spectrometry
DU Shu-lan
Abstract
DU Shu-lan
Abstract
A method for trace mercury in geochemical samples by cold vapor generation—atomic fluorescence spectrometry.The experimental conditions that carrying gas flow amount,Aqua Regina concentration and KBH4 concentration etc.The detection limit of the method for mercury is 0.023ug/L,linear scope is 0-4.0 ug/L,precision is 1.1-5.0(RSD%) for mercury.This method is demonstrated accurate and reliable through testing of standard samples.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A method for trace mercury in geochemical samples by cold vapor generation—atomic fluorescence spectrometry.The experimental conditions that carrying gas flow amount,Aqua Regina concentration and KBH4 concentration etc.The detection limit of the method for mercury is 0.023ug/L,linear scope is 0-4.0 ug/L,precision is 1.1-5.0(RSD%) for mercury.This method is demonstrated accurate and reliable through testing of standard samples.
Key concepts: Mercury (programming language), Detection limit, Chemistry, Cold vapour atomic fluorescence spectroscopy, Fluorescence spectrometry, Analytical Chemistry (journal), Mass spectrometry, Environmental chemistry