Determination of Glucose with Biosensor by Immobilization of Glucose Oxidase with Chitosan
Nan Cai
Abstract
Nan Cai
Abstract
A glucose biosensor was fabricated from a glucose oxidase-immobilized by chitosan and oxygen electrode.The effects of concentration of chitosan(0.3%),enzyme loading(0.8 mg),pH 7.0,phosphate buffer concentration(300 mmol/L),and temperature 25 ℃ for the response of the biosensor were investigated.The glucose biosensor has a linear response range of 0.016-1.10 mmol/L with a detection limit of 8.0 μmol/L(S/N=3).The response time was less than 60 s.The biosensor showed extremely good stability with a shelf-life of at least 3 months.The biosensor exhibited good repeatable response to a 0.25 mmol/L glucose solution with a relative standard deviation of 2.5%(n=10).The precision of fabrication of the biosensors using four different membranes was good with a RSD of 4.7%.Some common potential components in sample such as niacinamide,vitamin B6,vitamin B12,vitamin E,Ca2+,Mg2+,K+ and Zn2+ showed no interferences on the response of the glucose biosensor.The biosensor was successfully applied to determine the glucose in commercial beverage samples.
OpenAlex reports 4 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A glucose biosensor was fabricated from a glucose oxidase-immobilized by chitosan and oxygen electrode.The effects of concentration of chitosan(0.3%),enzyme loading(0.8 mg),pH 7.0,phosphate buffer concentration(300 mmol/L),and temperature 25 ℃ for the response of the biosensor were investigated.The glucose biosensor has a linear response range of 0.016-1.10 mmol/L with a detection limit of 8.0 μmol/L(S/N=3).The response time was less than 60 s.The biosensor showed extremely good stability with a shelf-life of at least 3 months.The biosensor exhibited good repeatable response to a 0.25 mmol/L glucose solution with a relative standard deviation of 2.5%(n=10).The precision of fabrication of the biosensors using four different membranes was good with a RSD of 4.7%.Some common potential components in sample such as niacinamide,vitamin B6,vitamin B12,vitamin E,Ca2+,Mg2+,K+ and Zn2+ showed no interferences on the response of the glucose biosensor.The biosensor was successfully applied to determine the glucose in commercial beverage samples.
Key concepts: Biosensor, Glucose oxidase, Chemistry, Chitosan, Chromatography, Detection limit, Clark electrode, Biochemistry